Inductance, Capacitance, Resistance Selection 

The 

L/C/R

 key selects the primary parameter measurement function. Each press of the key will 

select either inductance (

L

), capacitance (

C

) or resistance (

R

) along with the proper units of 

H

 

(henries), 

(farads) or 

 (ohms) in the main large display. 

Quality, Dissipation, Resistance Selection 

The 

Q/D/R

 key selects the secondary parameter measurement function. Each press of the key will 

select either the quality (

Q

) or dissipation (

D

) indicators or resistance (

) units in the small 

secondary display.  

Hold and Backlight Selection 

The 

HOLD      >2sec

 key selects the Hold feature and also enables the display backlight. 

Press the key and the 

H

 indicator will appear in the display and the last reading will “freeze” in the 

display. Press the key again and the reading will begin to update again. 
Press and Hold the key for 2 seconds and the display backlight will turn-on. To extinguish the 
backlight, press and hold the key again for 2 seconds or wait 1 minute for it to automatically disable. 

Minimum, Maximum and Average Selection 

The 

MAX MIN

 key selects the recording function. Press the key and the “

R

“ indicator will appear in 

the display and the meter will begin recording the minimum, maximum and average measured 
values. When this mode is entered, the auto power off and the function keys are disabled. 

Max-Min operation 

1.  Set all the function parameters for the test. 
2.  Press the MAX MIN key. The “R“ indicator will appear and a “beep” will sound after 

approximately six seconds. Two “beeps” will sound every time the max or min is updated. 

3.  Press the MAX MIN key. The “MAX” indicator and the maximum recorded value will appear in 

the display  

4.  Press the MAX MIN key. The “MIN” indicator and the minimum recorded value will appear in 

the display  

5.  Press the MAX MIN key. The “MAX - MIN” indicator and the difference between the maximum 

- minimum value will appear in the display  

6.  Press the MAX MIN key. The “AVG” indicator and the average of the recorded values will 

appear in the display. 

7.  Press and hold the MAX MIN key for 2 seconds to exit the mode. 

Notes

:  

The average value is a true average and averages up to 3000 values. If the 3000 limit is exceeded, 
the 

AVG

 indicator will flash and no further averaging will take place. The max and min values will 

continue to update. 
If the 

HOLD

 key is pressed during min max recording, the recording is halted until the 

HOLD

 key is 

pressed again. 

Relative Mode 

The relative mode displays the difference between the measured value and the value of a stored 
reference. 

1.  Press the REL key to enter the Relative mode. 
2.  The value in the display when the REL key is pressed will become the stored reference value 

and the display will indicate zero or a value close to zero (since the measured value and the 
reference value are the same at this point). 

3.  All subsequent measurements will be displayed as a value relative to the stored value. 
4.  The reference value may also be a value that was stored into memory using the SET Relative 

procedure (see the Setting a Relative Reference paragraph). 

5.  To use the SET relative value, press the SET key while in the Relative mode. 
6.  To exit the Relative mode, press and hold the REL key for 2 seconds. 

Содержание 380193

Страница 1: ...YVOLTAGETOTERMINALS DISCHARGE CAPACITOR BEFORETESTING AUTO POWER OFF REL Hi Lo LIMITS TOL 7 8 9 2sec HOLD MIN MAX SET 4 5 6 RANGE L C R Q D R 1 2 3 0 ENTER LCR Meter 380193 FREQ PAL SER User s Guide Passive Component LCR Meter Model 380193 ...

Страница 2: ...ated and with proper use will provide years of reliable service International Safety Symbols Caution Refer to the explanation in this Manual Caution Risk of electric shock Earth Ground Safety Precautions 1 Make sure any covers or battery doors are properly closed and secured 2 Always remove the test leads before replacing the battery or fuses 3 Inspect the condition of the test leads and the meter...

Страница 3: ...active n nano 10 9 H Data hold active micro 10 6 SET SET mode m milli 10 3 TOL Tolerance mode H Henry inductance units PAL Parallel equivalent circuit F Farad capacitance units SER Series equivalent circuit Ohms resistance units D Dissipation factor Upper limit Q Quality factor Lower limit R Resistance Relative mode L Inductance Low battery C Capacitance Tolerance percentage APO AUTO P AL c D 1kHz...

Страница 4: ...ntil APO OFF appears in the display Auto Power Off will also be disabled if the MIN MAX record mode is used or if the meter is powered by an external power supply Frequency Selection Press the FREQ key to select either 120Hz or 1kHz as the test frequency The frequency selected appears in the display Generally 120Hz will be used for large electrolytic capacitors and 1kHz used for most other tests P...

Страница 5: ...mately six seconds Two beeps will sound every time the max or min is updated 3 Press the MAX MIN key The MAX indicator and the maximum recorded value will appear in the display 4 Press the MAX MIN key The MIN indicator and the minimum recorded value will appear in the display 5 Press the MAX MIN key The MAX MIN indicator and the difference between the maximum minimum value will appear in the displ...

Страница 6: ...ings The selected will briefly appear in the small display 3 Previously stored user defined limits are accessed by pressing the SET key 4 The meter will sound an audible tone and blink the upper or lower limit indicator if the measured value is outside the limits 5 Press and hold the TOL key for 2 seconds to exit the mode Set Limits and Open Short Calibration Selection The SET key is used to 1 Set...

Страница 7: ...ator will flash and the previously stored reference will appear with the first digit flashing 2 To adjust the reference set the value of the flashing digit by pressing the appropriate numeric key The adjustment selection will then proceed through each digit from left to right 3 Press the ENTER key to store the value and continue on to the upper limit adjustment The upper limit indicator will flash...

Страница 8: ...cessTM and other database utilities USB cable part 421509 USBCBL Instructions for use of the PC interface are included on the supplied Program Disk and are beyond the scope of this operation manual For complete details and instructions refer to the HELP file on the supplied Program Disk You as the end user are legally bound EU Battery ordinance to return all used batteries disposal in the househol...

Страница 9: ...g 100 Cx 5d DF 0 1 after open cal 1999 9pF 1 0 rdg 5d DF 0 1 2 0 rdg 100 Cx 5d DF 0 1 after open cal Inductance 120Hz Range Lx accuracy DF 0 5 DF accuracy DF 0 5 Note 10000H Not specified Not specified 1999 9H 1 0 rdg Lx 10000 5d 2 0 rdg 100 Lx 5d after open cal 199 99H 0 7 rdg Lx 10000 5d 1 2 rdg 100 Lx 5d 19 999H 0 7 rdg Lx 10000 5d 1 2 rdg 100 Lx 5d 1999 9mH 0 7 rdg Lx 10000 5d 1 2 rdg 100 Lx 5...

Страница 10: ...r equivalent for high resistance calibration or certification Note In the 20 range effective readings must be over 20 counts Test Frequency accuracy 122 88Hz 4Hz and 1kHz 4Hz Display Dual 4 digit backlit LCD Overload indication OL Low battery indication Measurement rate One time per second Auto power off After 10 minutes of inactivity Operating environment 0o C to 50o C 32o F to 122o F 80 RH Stora...

Отзывы: