background image

380193-en-GB_V2.6   8/16 

6

Hi / Lo Limits Mode 

The Hi / Lo limits mode compares the measured value to the stored high and low limit values and 
gives an audible and visible indication if the measured value is outside the limits. See the setting 
Hi/Lo limits paragraph below to store the limits in memory. 

1. Press 

the 

Hi/Lo LIMITS

 key to enter the mode. The display will briefly show the stored upper 

limit with the “    “ indicator and then the stored lower limit with the “   “ indicator before 
displaying the measured value. 

2.  The meter will sound an audible tone and blink the upper or lower limit indicator if the 

measured value is outside the limits. 

3.  The meter will ignore an “OL” overload reading. 
4.  Press the Hi/Lo LIMITS key to exit the mode. 

% Tolerance Mode 

The % Tolerance Limits mode compares the measured value to a % high and low limit based on a 
stored reference value and gives an audible and visible indication if the measured value is outside 
the limits. Any % limit can be entered in the SET % Limit mode (see paragraph below) or standard 
1%, 5%, 10% and 20% symmetrical limits can be selected directly in the % Tolerance mode. 

1. Press 

the 

TOL

 key to enter the mode. The display will briefly show the stored reference value 

in the main display and the small display will indicate the % difference between the measured 
value and the reference value. See the SET % Limit paragraph to change the reference 
value. 

2.  Press the TOL key to step through and select the 1, 5, 10 or 20% settings. The selected % 

will briefly appear in the small display. 

3.  Previously stored user defined % limits are accessed by pressing the SET key. 
4.  The meter will sound an audible tone and blink the upper or lower limit indicator if the 

measured value is outside the limits. 

5.  Press and hold the TOL key for 2 seconds to exit the mode. 

Set Limits and Open/Short Calibration Selection 

The 

SET

 key is used to; 1. Set Hi/Lo limits, 2. Set % limits, 3. Set Tolerance reference value and 4. 

Perform the Open / Short calibration. The SET mode can only be activated if no other function is 
active. 

Entering the SET mode 

1.  Power ON and press the 

SET

 key.  

2.  The display will clear, “

SEt

“ will appear in the small display and a flashing     

TOL

 and flashing 

indicators will           appear in the display. 

3.  The 5 keys that are now active are; Power, SET, REL, Hi/Lo, and TOL 

Open and Short Calibration 

The Open and Short function removes stray parallel and series fixture impedances from the 
measured value. This feature improves accuracy for very high or low impedances. 
(Note: Remove any leads from the meter during this procedure. Leaving them connected will add 
impedance to the circuit causing the calibration to fail indicated by 

OUT UAL 

appearing on the 

display.) 

1. Press 

the 

SET

 key 2 times and the display will indicate “

CAL OPEn

”. 

2.  Remove any devices or test leads from the input terminals and press “ENTER” (PAL SER). 

After several seconds the calibration will complete and display “

CAL SHrt

”. 

3.  Short the input terminals and press “ENTER” (PAL SER). After several seconds the 

calibration will complete and the meter will return to normal operation. 

4.  Press the “SET” to bypass either the open or short calibration. 

Содержание 380193

Страница 1: ...YVOLTAGETOTERMINALS DISCHARGE CAPACITOR BEFORETESTING AUTO POWER OFF REL Hi Lo LIMITS TOL 7 8 9 2sec HOLD MIN MAX SET 4 5 6 RANGE L C R Q D R 1 2 3 0 ENTER LCR Meter 380193 FREQ PAL SER User s Guide Passive Component LCR Meter Model 380193 ...

Страница 2: ...ated and with proper use will provide years of reliable service International Safety Symbols Caution Refer to the explanation in this Manual Caution Risk of electric shock Earth Ground Safety Precautions 1 Make sure any covers or battery doors are properly closed and secured 2 Always remove the test leads before replacing the battery or fuses 3 Inspect the condition of the test leads and the meter...

Страница 3: ...active n nano 10 9 H Data hold active micro 10 6 SET SET mode m milli 10 3 TOL Tolerance mode H Henry inductance units PAL Parallel equivalent circuit F Farad capacitance units SER Series equivalent circuit Ohms resistance units D Dissipation factor Upper limit Q Quality factor Lower limit R Resistance Relative mode L Inductance Low battery C Capacitance Tolerance percentage APO AUTO P AL c D 1kHz...

Страница 4: ...ntil APO OFF appears in the display Auto Power Off will also be disabled if the MIN MAX record mode is used or if the meter is powered by an external power supply Frequency Selection Press the FREQ key to select either 120Hz or 1kHz as the test frequency The frequency selected appears in the display Generally 120Hz will be used for large electrolytic capacitors and 1kHz used for most other tests P...

Страница 5: ...mately six seconds Two beeps will sound every time the max or min is updated 3 Press the MAX MIN key The MAX indicator and the maximum recorded value will appear in the display 4 Press the MAX MIN key The MIN indicator and the minimum recorded value will appear in the display 5 Press the MAX MIN key The MAX MIN indicator and the difference between the maximum minimum value will appear in the displ...

Страница 6: ...ings The selected will briefly appear in the small display 3 Previously stored user defined limits are accessed by pressing the SET key 4 The meter will sound an audible tone and blink the upper or lower limit indicator if the measured value is outside the limits 5 Press and hold the TOL key for 2 seconds to exit the mode Set Limits and Open Short Calibration Selection The SET key is used to 1 Set...

Страница 7: ...ator will flash and the previously stored reference will appear with the first digit flashing 2 To adjust the reference set the value of the flashing digit by pressing the appropriate numeric key The adjustment selection will then proceed through each digit from left to right 3 Press the ENTER key to store the value and continue on to the upper limit adjustment The upper limit indicator will flash...

Страница 8: ...cessTM and other database utilities USB cable part 421509 USBCBL Instructions for use of the PC interface are included on the supplied Program Disk and are beyond the scope of this operation manual For complete details and instructions refer to the HELP file on the supplied Program Disk You as the end user are legally bound EU Battery ordinance to return all used batteries disposal in the househol...

Страница 9: ...g 100 Cx 5d DF 0 1 after open cal 1999 9pF 1 0 rdg 5d DF 0 1 2 0 rdg 100 Cx 5d DF 0 1 after open cal Inductance 120Hz Range Lx accuracy DF 0 5 DF accuracy DF 0 5 Note 10000H Not specified Not specified 1999 9H 1 0 rdg Lx 10000 5d 2 0 rdg 100 Lx 5d after open cal 199 99H 0 7 rdg Lx 10000 5d 1 2 rdg 100 Lx 5d 19 999H 0 7 rdg Lx 10000 5d 1 2 rdg 100 Lx 5d 1999 9mH 0 7 rdg Lx 10000 5d 1 2 rdg 100 Lx 5...

Страница 10: ...r equivalent for high resistance calibration or certification Note In the 20 range effective readings must be over 20 counts Test Frequency accuracy 122 88Hz 4Hz and 1kHz 4Hz Display Dual 4 digit backlit LCD Overload indication OL Low battery indication Measurement rate One time per second Auto power off After 10 minutes of inactivity Operating environment 0o C to 50o C 32o F to 122o F 80 RH Stora...

Отзывы: