Measuring Power
48
IQS-1700
Measuring Corrected Power
To delete a user reference from the list:
1.
From the
Settings
function tab, select the
Lists
tab.
2.
Select the value to remove by clicking it once.
3.
Click
to remove the value.
Measuring Corrected Power
Applying a correction factor to the measured power is useful when
compensating for known inaccuracies (power gains or losses) at specific
wavelengths.
A correction factor (CF) can be applied to any measurement that is
displayed in either dB, dBm, W, or W/W.
When a dB correction factor different from 0.000 is defined, the displayed
power is equal to the actual power plus the value of the correction factor.
When a W/W correction factor different from 1.000 is defined, the
displayed power is equal to the actual power times the value of the
correction factor.
Note:
With an active correction factor, the CF marker appears in the data display
for the matching channel.
Note:
Special care must be taken when setting a CF as it emulates a new
calibration (the module calibration is not affected and will be restored by
setting back the CF to a null value).
When expressed in dB, the CF can be a positive or negative value. When
the currently selected measurement unit is W, the correction factor is
expressed in W/W, indicating a multiplication factor for the current
wavelength and channel. The CF expressed in W/W will always be a
positive value.
Содержание IQS-1700
Страница 1: ...User Guide High Performance Power Meter and Optical Head for IQS 600 IQS 1700...
Страница 5: ...Contents High Performance Power Meter v Index 189...
Страница 8: ......
Страница 14: ......
Страница 16: ......
Страница 44: ......
Страница 64: ......
Страница 80: ...Performing Acquisitions 72 IQS 1700 Consulting Acquired Data To view your acquisition data select the Data tab...
Страница 96: ......
Страница 114: ......
Страница 196: ......