SCPI Command Reference
High Performance Power Meter
117
Product-Specific Commands—Description
:INITiate[1..n]:AUTO?
Description
This query returns a value indicating whether a
programmed (or "Autostop") acquisition is in
progress.
Syntax
:INITiate[1..n]:AUTO?
Parameter(s)
None
Response Syntax
<AcqOnOff>
Response(s)
AcqOnOff:
The response data syntax for <AcqOnOff> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.
This query returns the current <AcqOnOff>
acquisition state:
0 - Autostop acquisition is stopped.
1 - Autostop acquisition is running.
Example(s)
INIT:AUTO?
See Also
ABORt
INITiate:AUTO
INITiate:EXTRema
INITiate:EXTRema?
MMEMory:ACQuisition
MMEMory:ACQuisition?
TRACe:POINt