SCPI Command Reference
High Performance Power Meter
137
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:COUNt
Description
Sets the number of measurements used to
calculate the final measurement's average on the
specified channel.
At *RST, this value is 2.
Syntax
:SENSe[1..n]:AVERage:COUNt<wsp><AverageC
ount>|MAXimum|MINimum|DEFault
Parameter(s)
AverageCount:
The program data syntax for <AverageCount> is
defined as a <numeric_value> element. The
<AverageCount> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <AverageCount> parameter.
The <numeric_value> represents the number of
measures to average for the final measures.
MIN, MAX and DEF can also be used as
parameters.
Example(s)
SENS:AVER:COUN 12
See Also
SENSe:AVERage:COUNt?
SENSe:AVERage:STATe
SENSe:AVERage:STATe?