Maintenance
Variable Attenuator/Variable Reflector
87
Cleaning Connectors Equipped with EUI/EUA Adapters
3.
Gently wipe the connector and ferrule.
I
MPORTANT
Isopropyl alcohol takes approximately ten seconds to evaporate.
Since isopropyl alcohol is not absolutely pure, evaporation will leave
microscopic residues. Make sure you dry the surfaces before
evaporation occurs.
4.
With a dry lint-free wiping cloth, gently wipe the same surfaces in
order to be sure that the connector and ferrule are perfectly dry.
5.
Throw out the wiping cloths after one use.
W
ARNING
DO NOT verify the surface of the connector with a fiber-optic
microscope WHILE THE UNIT IS ACTIVE. Doing so WILL result in
permanent eye damage.
6.
After having cleaned the connectors and having made sure that the
unit is not active, verify the surface of the connector with a small,
portable fiber-optic microscope.