IQ-3100 and IQ-3300 Remote Control
76
IQ-3100/IQ-3300
IQ-3100/IQ-3300 Specific Commands
OFFSET?
Description
This query returns the offset value.
Syntax
OFFSET(0..26)?
Response
The current offset is in the “±99.999” format. The units
are dB.
Example
OFFSET3?
OUTPut:APMode
Description
This command selects absolute or relative attenuation or
reflection display. When “absolute” is selected, the absolute
attenuation or reflection introduced by the module is displayed.
When “reference” is selected, attenuation or reflection relative
to the current reference is displayed. When “X+B” is selected,
the attenuation or reflection is modified by the correction factor.
Syntax
OUTP(0..26):APM<space><value>
Parameters
The <value> parameter can be
“0” Relative display mode
“1” Absolute display mode
“2” X+B display mode
Example
OUTP3:APM 1