10
CPMA a n d CPMB
Guidance and manufacturer’s declaration – electromagnetic immunity (table 2 according IEC 60601-1-2)
The telescopic pillars CPMA and CPMB are intended for use in the electromagnetic environment specified below.
The customer or user of the telescopic pillars should ensure that they are used in such an environment.
IMMUNITY test
IEC 60601
test level
Compliance level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at
least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Mains power quality should be that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U1
(>95% dip in U1)
for 0.5 cycle
40% U1
(60% dip in U1)
for 5 cycles
70% U1
(30% dip in U1)
for 25 cycles
<5% U1
(>95% dip in U1)
for 5 s
<5% U1
(>95% dip in U1)
for 0,5 cycle
40% U1
(60% dip in U1)
for 5 cycles
70% U1
(30% dip in U1)
for 25 cycles
<5% U1
(>95% dip in U1)
for 5 s
Mains power quality should be that of a typical commercial or
hospital environment. If the user of telescopic pillars CPMA and
CPMB requires continued operation during mains power
interruptions, it is recommended that the telescopic pillars CPMA
and CPMB are powered from an uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
If distortion occurs, it may be necessary to position the telescopic
pillars CPMA and CPMB further from sources of power frequency
magnetic fields, or to install magnetic shielding. The power
frequency magnetic field should be measured in the intended
installation location to ensure that it is sufficiently low.
NOTE: U1 is the a.c. mains voltage prior to application of the test level.