10. Attach short length’s of ¼” and 3/8” OD tubing to the TO flush valve and product water 3-
way sample valve connection located near the TO flush valve.
11. Refer to figure 9. Locate the 10-foot power cord attached to the control enclosure. Connect
the fork connectors on this cord to the terminals on the 24 VAC transformer. Systems not
utilizing the booster pump require no additional wiring. Systems utilizing the booster pump
require the addition of the Y-connector included in the small parts kit to enable connecting
the electrical connection on the booster pump to the control enclosure circuit.
Do not connect the transformer to the 115-Volt receptacle at this time.
Figure 9
Figure 7
Figure 8
INLET WATER SUPPLY
50 PSIG MAX.
TO 120 VAC OUTLET
A
B
C
SOLENOID
VALVE
1
5
2
4 3
IN
A
B
C
TO 120 VAC OUTLET
SOLENOID
VALVE
IN
(COVER REMOVED)
3
4
2
5
1
CONTROL ENCLOSURE
(COVER REMOVED)
CONTROL ENCLOSURE
TRANSFORMER
LEVEL
SWITCH
LEVEL
SWITCH
TRANSFORMER
BOOSTER PUMP
Y-HARNESS
(SHIPPED LOOSE)
ELECTRICAL INTERCONNECTING DIAGRAM
WITHOUT
BOOSTER PUMP
WITH
BOOSTER PUMP
Figure 6
Transformer shown (75
VA capacity) is for
reference only.
Some models include a
50 VA capacity
receptacle mounted
version. In those
applications, connect
the forked terminals to
the two screw terminals
as required.
Either transformer can
be used in this
application.
Transformer shown (75
VA capacity) is typical
for systems with a
booster pump.
50 VA capacity
receptacle mounted
version is
not
suitable
for this application.