background image

 

16 

13. Rotate the handle on the product water 3-way sample valve so it is pointing towards 

the storage tank connection (rear of unit). 

 

Performance Monitoring (optional)

 

 
Once  the  system  has  been  flushed  initially,  the  outlet  and  drain  water  flow  rates  and  water 
quality  can  be  measured.  These measurements  can  be  taken  at  any  time  and  compared  with 
the  initial  values  to  detect  any  changes  in  system  performance.  Product  water  volume  and/or 
quality are affected by a number of factors including inlet water pH, dissolved solids, pressure 
and temperature. Changes in these parameters will affect system performance. 
 
Published performance specifications are based on certain standardized conditions. Variations 
from  these  conditions  can  be  expected  to  affect  actual  measured  performance.  These 
conditions assume new filter/RO cartridges, differential pressure across the RO module equal to 
specification  pressure  and  inlet  water  temperature  at  the  specification  value.  Increases 
(decreases)  in  pressure  and/or  temperature  will  result  in  increased  (decreased)  production 
rates. 
 
Close the tank inlet valve and open the product water

 

sample valve to obtain a sample. The flow 

rates can be measured with a stopwatch and a graduated cylinder or calibrated measuring cup. 
The drain water flow rate can be measured from the end of the red tubing in a similar manner.   
 
 

Inlet water flow = Outlet (Product) Water Flow + Drain Water Flow 

 
Water  quality  can  be  determined  by  measuring  the  total  dissolved  solids  (TDS)  of  the  outlet 
water  sample.  A  laboratory  or  hand  held  TDS/conductivity  tester  should  be  used  for  this 
measurement.  Optionally,  TDS  measurements  are  used  to  calculate  Total  Dissolved  Solids 
percent  reduction  as  an  indication  of  system  performance.  Percent  reduction  can  be  used  to 
determine when the RO cartridge should be replaced. 
 

TDS Measurement 

 

1.  Rinse a clean cup or graduated cylinder twice with the outlet water. Then collect enough 

fresh sample to measure with your conductivity tester. 

 
2.  Following  the  instructions  of  your  conductivity  tester,  measure  and  record  the 

TDS/conductivity value of the outlet water sample.   

 

 

 

TDS Reduction Calculation (optional):

 

 

1.  Obtain an outlet water sample and TDS measurement as described above. 
 
2.  Obtain  an  inlet  water  sample  in  the  same  manner  as  above,  but  from  the  inlet  water 

source to the MRS-200 system. 

 
3.  Using  your  TDS/conductivity  tester,  measure  and  record  the  TDS/conductivity  value  of 

the inlet water sample. 

 
4.  To  determine  percent reduction,  apply  the  two TDS/conductivity  values  obtained  to  the 

following equation: 

 

 

 

Inlet TDS – Outlet TDS 

Inlet TDS 

x 100 = % TDS Reduction

 

 

Содержание MRS-200

Страница 1: ...1 Cat No EV311117 Rev B 11 6 03 DCO 7060 MODEL MRS 200 Reverse Osmosis Mineral Reduction System Installation and Start up Guide...

Страница 2: ...Installation 4 Cartridge Installation and Activation 10 Start Up and Blend Adjustment 10 Adjusting the Recovery 12 Operation 14 Maintenance 14 Performance Monitoring 16 TDS Measurement 16 Performance...

Страница 3: ...hlorine An activated carbon filter has been provided with this system to protect the reverse osmosis membrane from chlorine attack Influent chlorine should not exceed 3 mg L The MRS 200 System consist...

Страница 4: ...nd permit access for maintenance The system will weigh approximately 175 pounds when the storage tank is full 4 ____ Does the inlet water supply meet the requirements listed below 1 A minimum of 1 0 G...

Страница 5: ...nts Place the main unit on the shelf positioning the cartridge heads directly over the cutout area Insert a cable tie into one of the holes in the baseplate and through the wire slats in the shelf mak...

Страница 6: ...A MP S 2 INS E RT E LB OW FULLY 3 S LIDE LOCK ING CLIP UP TO S E CURE 1 S LIDE LOCK ING CLIP DOW N Figure 3 4 Water supplies with a dynamic operating pressure of 50 PSI or less require the use of the...

Страница 7: ...UMP 10 NUT 8 1 2 X 1 SPACER 4 10 X 1 1 2 SCREW 8 10 WASHER 16 MOUNTING PLATE Figure 4 5 Refer to Figure 5 for a sample general arrangement Use this as a guideline for determining valve location tubing...

Страница 8: ...ow control insert red plastic cylinder in small parts kit Press this insert into the red OD tubing Fig 6 Attach the end of the tubing with the flow control insert into the 3 8 stem x OD tube reducing...

Страница 9: ...lt receptacle at this time Figure 9 Figure 7 Figure 8 INLET WATER SUPPLY 50 PSIG MAX TO 120 VAC OUTLET A B C SOLENOID VALVE 1 5 2 4 3 IN A B C TO 120 VAC OUTLET SOLENOID VALVE IN COVER REMOVED 3 4 2 5...

Страница 10: ...ative solution from the RO cartridge and should not be consumed 6 If you are utilizing the capillary tubing assembly refer to the section titled Adjusting the Recovery for instructions Return to this...

Страница 11: ...to contain a range of TDS greater than what an RO system normally produces In those cases a filtered water blend valve has been provided to blend in filtered not RO treated water to attain a specific...

Страница 12: ...product flow with a 1000 ml graduated cylinder Don t worry if the flow is slightly more or less than what is specified 2 Determine the appropriate concentrate flow for your system and the desired rec...

Страница 13: ...480 70 520 80 554 10 00 50 408 60 450 70 489 80 521 12 00 50 363 60 402 70 437 80 467 14 00 50 330 60 366 70 399 80 427 16 00 50 307 60 342 70 371 80 397 18 00 50 292 60 326 70 352 80 377 20 00 50 28...

Страница 14: ...or 6 000 12 000 18 000 or 24 000 gallons d Calendar scale is jumper selectable for 6 months or 12 months Monitor operation a Active green light blinks every two seconds during any minimum water flow c...

Страница 15: ...s or until the water is clear 10 Close the TO flush valve 11 Connect the 24 VAC Transformer from the 115 Volt receptacle Every 6 12 months Sanitize the storage tank 1 Refer to Start up section steps 1...

Страница 16: ...and a graduated cylinder or calibrated measuring cup The drain water flow rate can be measured from the end of the red tubing in a similar manner Inlet water flow Outlet Product Water Flow Drain Wate...

Страница 17: ...has removed 92 of the dissolved solids from the inlet water 6 The RO cartridge should be changed every 1 to 1 1 2 years or when the percent reduction falls below 90 A form has been provided below to r...

Страница 18: ...valve as required to attain specified water quality a Tubing piping run to equipment too restrictive a Increase tubing piping diameter Refer to System Installation section for recommendations 4 Low w...

Страница 19: ...19 Replacement Parts...

Страница 20: ...EV307850 Support Ring EA 2 49 EV310445 Label Carbon Filter EA 1 50 EV310446 Label RO Module EA 1 51 01010248 Adaptor 3 8 Stem X 1 4 Tube EA 1 53 EV306996 Directional Arrow EA 1 No Part Number Descrip...

Страница 21: ...hole punched Prints black ink on 50 offset white VENDOR MUST SUBMIT A MATERIAL CERTIFICATE OF COMPLIANCE WITH EACH SHIPMENT LET CHANGE BY APRVD DATE A DCO 4785 RH SK 9 12 02 B DCO 7060 KR RH 11 6 03 T...

Отзывы: