Instruction Manual
CON 11/110
66
11 ERROR
MESSAGES
The following table provides a guideline to enable diagnosis of possible problems
indicated by the messages generated by the meter. The table also provides possible
solutions to the problems encountered.
ERROR MESSAGE
INDICATES
POSSIBLE CAUSE
CORRECTIVE
ACTION
ERR. Annunciator
Wrong keypad input
Wrong input in
selected mode.
Release key. Select
valid operations
depending on mode.
Secondary Display
Toggles Continuously
between Calibration
Standard Values
Calibration error
Incorrect calibration
solution used or
contamination of
calibration solution
occurring in automatic
calibration mode
Check if the right
calibration solution
was selected or use
fresh calibration
solution.
Battery icon lights up
Low Battery
Battery power is low
Replace batteries
with fresh ones as