Instruction Manual
CON 11/110
74
18
CALCULATING TEMPERATURE COEFFICIENTS
To determine the temperature coefficient of your sample solution use this formula:
Where:
tc
= Temperature coefficient
25
= 25 °C
C
T1
= Conductivity at Temp 1
C
T2
= Conductivity at Temp 2
T
1
= Temp 1
T
2
= Temp 2
NOTE:
A controlled temperature water bath is ideal for this procedure
1. Immerse the electrode into a sample of your solution and adjust the temperature
coefficient to 0% (that is, no compensation) by performing the following:
2. From measurement mode, press the
SETUP
key to enter Setup menu.
3. Press the
MI/
S
or
MR/
T
key until the lower display reads P7.0.
4. Press the
HOLD/ENTER
key to display the temperature coefficient value.
5. Press the
MR/
T
key until the upper display shows 0.0.
6. Press
HOLD
/
ENTER
key to confirm the value.
7. Press
CAL/MEAS
key to return to measurement mode.
8. Wait for 5 minutes. Note
T
1
and
C
T1
(conductivity at
T
1
).
9. Condition the sample solution and electrode to a temperature (
T
2
) that is about 5
°C to 10 °C different from
T
1
, and note the conductivity reading
C
T2
.
NOTE:
Record your results for future reference. Ideally
T
1
and
T
2
should bracket your
measurement temperature, and should not differ by more than 5 °C.
10. Calculate the temperature coefficient of your solution using the formula shown
above.
11. Enter the temperature coefficient you calculated into the meter by following the
following steps.
12. From measurement mode, press the
SETUP
key to enter Setup mode.
13. Press the
MR/
T
key until the lower display reads P7.0.
14. Press the
HOLD/ENTER
key to show the temperature coefficient value (should be
0, as per step 1 above).
15. Press the
MI/
S
key until the upper display shows your calculated temperature
coefficient.
16. Press
HOLD
/
ENTER
key to confirm the value.
17. Press
CAL/MEAS
key twice to return to measurement mode.