Instruction Manual
CON 11/110
45
8.6
P5.0: Reset to factory defaults
Program 5.0 reset functions lets you have an option to only reset all the meter’s
calibrations and its data that have been stored or reset all calibrations, its stored
data, memory, and any other setup functions that you might have changed back to
factory default settings. See Factory Default Settings at page 69 and 70.
8.6.1 P5.1: Calibration Reset
This reset function clears all the calibrations and its data that
have been set by you. All other setup parameters that you
have changed are not reset.
From measurement mode
1. Press SETUP key to enter Set Up menu.
2. Press the MI/
▲
or MR/
▼
keys to until you view
parameter P5.0.
3. Press the HOLD/ENTER key to enter parameter
P5.1 with the upper display showing “CAL”.
4. Press the HOLD/ENTER key again to enter the
calibration reset selection mode.
5. Press the MI/
▲
or MR/
▼
keys to toggle between NO
and YES.
•
NO retains current settings
•
YES clears all calibrations and its data.
6. Press the HOLD/ENTER key to confirm selection
and to return to measurement mode. Otherwise
press CAL/MEAS key to return to measurement
mode without resetting the calibration data.
SETUP
cal
p 5.1
SETUP
cal
SETUP
SETUP
yes
MI
MR
HOLD
ENTER
HOLD
ENTER
p 5.0
cal
Figure 38: P5.1-Clear all
calibration and its data