Instruction Manual
CON 11/110
48
8.8 P7.0:
Temperature
8.8.1 P7.1:
Setting the temperature coefficient
The temperature coefficient is the amount of change in conductivity per degree of
temperature; it is expressed in percent per °C.. Entering the exact temperature
coefficient of your solution lets you accurately compensate
temperature for almost any solution. You can adjust 0.0 to
10.0 % per °C. Meter default is 2.1% per °C. By setting the
temperature coefficient to 0.0%, the meter will not apply any
compensation but will display actual temperature.
From measurement mode
1. Press SETUP key to enter Set Up menu.
2. Press the MI/
▲
or MR/
▼
keys until you view
parameter P7.0.
3. Press the HOLD/ENTER key to select parameter
7.1. The display shows “tCO” on the upper display.
4. Press the HOLD/ENTER key again. The upper
display shows the default temperature coefficient
“2.1”.
5. Press the MI/
▲
or MR/
▼
keys to select the
temperature coefficient of your solution.
6. Press the HOLD/ENTER key to confirm.
7. Press the CAL/MEAS key twice to return to the
measurement mode.
Notes
If you do not know the temperature coefficient of your
solution you can determine the correct value using the
formula in page 74 “Calculating Temperature Coefficients”.