Instruction Manual
CON 11/110
21
5.2
Manual Temperature Compensation
To select the Manual Temperature Compensation (MTC) option, set up the Manual
Temperature Compensation mode in the SETUP menu Program P 1.3 by setting to
‘NO”.
1. Press SETUP key to enter Set Up mode. The display shows “COF” at “P
1.0”.
2.
Press the HOLD/ENTER key to scroll through subgroups until you view
“ATC” at “P1.3” with small annunciators of “°C” and “ATC” or “°F” and “ATC”.
3. Press the HOLD/ENTER key again. The upper display shows “ATC” and the
lower display shows “YES”.
4. Press
the
MI/
▲
or MR/
▼
key to select the lower display to show “NO”.
5. Press the HOLD/ENTER key to confirm selection and to return to the
subgroup menu.
6. Press the CAL/MEAS key to return to measurement mode.