Operating Instructions
α
lpha-CON1000
68X216802
7
3.2
Back Panel
The back panel consists of two connectors. The first connector is the 17-way PCB edge connector and
the other is the 5-way connector.
Connection for the 17-way screw terminals (from left to right):
1. AC mains live wire
10. Alarm relay resting position (NO)
2. AC mains neutral wire
11. Alarm relay common
3. AC mains protective earth wire
12. Alarm relay working position (NC)
4. Low set relay resting position (NC)
13. Hold function switch terminal 1
5. Low set relay common
14. Hold function switch terminal 2
6. Low set relay working position (NO)
15. No connection
7. High set relay resting position (NC)
16. 0/4 - 20 mA for -ve connection
8. High set relay common
17. 0/4 - 20 mA for +ve connection
9. High set relay working position (NO)
Connections for the 5-way screw terminals:
18. Pt1000/Pt100 lead 1 terminal
19.
Pt1000/Pt100 sense lead terminal
20.
Pt1000/Pt100 lead 2 terminal
21. Conductivity lead 1
22. Conductivity lead 2
IMPORTANT: The Alarm relay functions as an “Active Low” device i.e. it switches OFF under
Alarm condition. Therefore the Alarm display device should be connected to the ‘NC’ contacts of
the relay.
Pt100/
RELAY1
L
N
PE
63mA (F)
FUSE 250VAC
HOLD
ALARM
RELAY2
Pt1000
NC
-
+
cell
J2