Operating Instructions
α
lpha-CON1000
68X216802
23
1)
Follow directions in 7.9.2 to enter Configuration mode.
2)
Press the ENTER key. Scroll with the
∆∆
or
∇
∇
keys until the upper
display shows “Stdy” or “FLEt” and lower display shows “AL.C.”.
-
AL.C = alarm contact
-
StdY = steady contact
-
FLEt = fleeting (single pulse) contact
3)
Press the
∆∆
or
∇
∇
keys to select steady or pulse contact.
4)
Press the ENTER key to confirm your selection.
5)
Proceed to 7.9.4, or return to Measurement mode by pressing the
∆∆
and
∇
∇
keys simultaneously (escape).
7.7.4
Input Line Resistance Adjust
This function compensates for the line resistance of the cable to its cell.
1)
Follow directions in 7.9.3 to enter Configuration mode.
2)
Press the ENTER key. Scroll with the
∆∆
or
∇
∇
keys until the
upper display shows “0.0” and lower display shows “L.Ad”.
-
L.Ad = line adjuster resistance (0.0 to 100.0)
3)
Press the
∆∆
or
∇
∇
keys to input the value.
4)
Press the ENTER key to confirm your selection.
5)
Proceed to 7.9.5, or return to Measurement mode by pressing the
∆∆
and
∇
∇
keys simultaneously
(escape).
7.7.5
Reverting to factory default settings
Use this parameter to reset all settings to factory default. Changing from “no” to “YES” and pressing the
ENTER key resets all settings to factory default.
WARNING: If you select yes, all the settings you have made will be
overwritten as a result!
1)
Follow directions in 7.9.4 to enter Configuration mode.
2)
Press the ENTER key. Scroll with the
∆∆
or
∇
∇
keys until the upper
display shows “no” or “YES”, and the lower display shows “deF”
(default).
3)
Press the
∆∆
or
∇
∇
keys to select no or yes. Selecting “YES” and pressing the ENTER key will
overwrite all setting you have made and automatically return you to Measurement mode.
4)
Press the ENTER key to confirm your selection and to return to Advanced Set-up mode, or return
to Measurement mode by pressing the
∆∆
and
∇
∇
keys simultaneously (escape).
7.8
Calibration (CAL) sub-function
The calibration procedure in Advanced Set-up mode is identical to the procedure in the Calibration
mode. The only difference is that the controller will revert back to Set-up mode (instead of Measurement
mode) after calibration is completed.
no
no
dEF
dEF
HOLD SETUP
0.0
0.0
L.Ad
L.Ad
HOLD SETUP
Stdy
Stdy
AL.C
AL.C
HOLD SETUP
FLEt
FLEt
AL.C
AL.C
HOLD SETUP