68X216802
7.4.3
Choosing break/make contact relay type
Note: If the controller type “oFF” is set, the parameters listed in 7.5.3, 7.5.4, 7.5.5 and 7.5.6 are blanked
out.
This mode lets you determine the relay-state under Non-Alarm condition – dEEN (de-energised) or EN
(energised).
1)
Follow directions in 7.6.2 to enter Controller mode.
2)
Press the ENTER key. Scroll until the lower display shows “rEL”
and the upper display shows the current selection (de-energised =
dEEN or energised = EN).
3)
Press the
∆∆
or
∇
∇
keys to choose de-energised or energised
relay state.
4)
Press the ENTER key to confirm your selection.
5)
Continue with Advanced Set-up mode procedures, or return to
Measurement mode by pressing the
∆∆
and
∇
∇
keys simultaneously
(escape).
7.5
Measurement Range sub-function
In this sub-function, the appropriate range is selected with the appropriate cell constant.
7.5.1
Selecting the Measuring Range sub-function
1)
Enter Advanced Set-up mode. Push the ENTER key and scroll to
Advanced Set-up security code “22”. Push the ENTER key again.
2)
Press the
∆∆
or
∇
∇
keys to scroll until the upper display shows
“rng”.
3)
Press the ENTER key. The upper display shows the maximum
measurement range, while the lower display shows the Cell constant. In
the lower left-corner of the LCD, the number corresponding to the
respective range is displayed.
4)
Press the
∆∆
or
∇
∇
keys to select the correct range. (Please refer to
Appendix 2 for the full list of measurement ranges).
5)
Proceed to 7.8 to set current output or
return to Measurement mode by pressing the
∆∆
and
∇
∇
keys simultaneously (escape).
dEEn
dEEn
rEL
rEL
HOLD SETUP
En
En
rEL
rEL
HOLD SETUP
rng
rng
HOLD SETUP
200
200..00
0.1
0.1
HOLD SETUP
4
µ
S