Manufacturer
Distributor
18
9.
Press the button
Calibrate
.
Figure 21. Calibration Process
10.
Wait until the application displays a notification window with the message
Disconnect the probe from
the central tube
(Fig. 22).
Figure 22. Disconnecting the Probe
11.
Disconnect the probe and wait until the application displays the message
Done!
showing completion
of the calibration process (Fig. 23).
Figure 23. Completion of Calibration
IMPORTANT NOTE:
Should an error arise during calibration, repeat the process anew.
As a result of calibration, the voltage provided by the sensor will be 0 V when dry and approximately 9.8 V when
completely filled with fuel.