Display of LED Error
Instruction Manual ATS 400 with SC-NG
Version 4.03
45
9.2.3
Overtemperature
Error number: 3
Exceeds the temperature in the device 70° C this situation will be triggered. If the
temperature drops below 55° C again, the error disappears.
Ensure a sufficiently low ambient temperature.
9.2.4
Loss of communication with the HMP
Error number: 4
If this code remains after power off and on of the device a board is defect and the
device must be sent in for repair.
9.2.5
Loss of communication with the MEP function
Error number: 5
If this code remains after power off and on of the device a board is defect and the
device must be sent in for repair.
9.2.6
Loss of communication with the MEP insulation test
Error number: 6
If this code remains after power off and on of the device a board is defect and the
device must be sent in for repair.
9.2.7
Loss of communication with the MEP protective earth test
Error number: 7
If this code remains after power off and on of the device a board is defect and the
device must be sent in for repair.