It is a 2-speed fan and at lower set point temperatures the fan speed 
will increase (shown by indicator 02 on the controller).
Allow at least 150 mm of space all around the unit and DO NOT obstruct 
any of the ventilation holes.

STAND

The stand can be folded flat against the under side of the unit when not 
in use or, if the unit is to be operated in a horizontal position, it can be 
swung down and forward until it hits the stops so that the unit can be 
used in an inclined position. The stand was not designed to be a carrying 
handle and should NOT be forced past the stops for this purpose.

DRY-WELL

The aluminium dry-well is located on the front panel. The dry-well is 
designed to accept two inserts: Ø4.8 and 6.4 mm as standard.  Other 
inserts are available to suit standard probe diameters. Use the nearest, 
larger size insert to the probe diameter being checked.  The inserts are 
a close fit in the dry-well to give good thermal conduction. Keep the 
inserts clean and avoid damage by storing carefully. The inserts MUST 
be regularly removed and cleaned to ensure they do not seize in the 
dry-well. DO NOT introduce any liquids or substance into the dry-well 
or inserts as this may result in the inserts lodging in the dry-well. 

TEMPERATURE CONTROLLER

In normal operating mode the numeric LED display shows the actual 
well temperature.  A Control Output Indicator light (O1) is located in 
the top left corner of the display. This indicates when the unit is in heat 
mode or when the unit is controlling within the proportional control band.
To show the set point temperature in the display, press and release 

either

 the up or down arrow button (

˄˅

). Press and hold 

either

 the up 

or down arrow button (

˄˅

) to change the set-point value.

A shift can be entered into the controller by pressing the scroll (left-hand) 
button.  Press and hold either the up or down arrow button (

˄˅

) to 

change the shift value. A factory set value will be entered at the time of 
certification. The shift value should only be adjusted if the unit is being 
monitored by an ETI Reference Thermometer or similar.
All other functions of the controller have been factory set and locked to 
maintain accuracy and repeatability. 

Содержание 271-401

Страница 1: ...3101 Dry Well heat cool source calibrator User Manual order code 271 401...

Страница 2: ...2...

Страница 3: ...4 safety 4 parts control 5 power lead 5 power switch 5 fan 6 stand 6 dry well 6 temperature controller 6 operation 7 temperature setting 7 probe testing 7 maintenance 8 care of the 3101 Dry Well 8 war...

Страница 4: ...pon demand The required temperature may be adjusted in 0 1 C increments across the range of the instrument using the buttons on the front panel These units are ideal for use in a variety of industrial...

Страница 5: ...connected to grounded AC power supply 100 240VAC 50 60Hz The unit requires 160 watts of power Mains plug fuse rating is 3 amps The 3101 dry well heat source is programmed and calibrated at the factor...

Страница 6: ...storing carefully The inserts MUST be regularly removed and cleaned to ensure they do not seize in the dry well DO NOT introduce any liquids or substance into the dry well or inserts as this may resul...

Страница 7: ...required to reach a higher set point temperature It will take longer to reach a lower set point temperature Once the set point temperature is reached a further 10 minutes should be allowed for stabili...

Страница 8: ...ire service or repair please contact ETI for assistance There are no user serviceable parts inside and any attempted repair will void any guarantee GUARANTEE This instrument carries a one year guarant...

Страница 9: ...r 10 minutes stabilisation 5 minutes cooling times ambient to 0 C after 10 minutes hole dimensions 2 x 13 mm that will accept various inserts inserts each unit is supplied with two inserts of customer...

Страница 10: ...ated against reference instrument s calibrated by a UKAS Accredited Calibration Laboratory at 10 C 0 C and 100 C To BS EN 60584 INSTRUMENT RECORD CARD 3101 Dry Well order code 271 401 manufactured by...

Отзывы: