9
LS720 Microscope Manual 2021
1.
Click on the "Compatibility" tab
2.
Click the "Change high DPI settings" button near the bottom of the menu
3.
Check the "Override high DPI scaling behavior" box and then select "System (Enhanced)"
right below it
4.
Click "OK" to close the menus
Restart Lumaview.
C.
Initialization and Calibration
1.
The LS720 automatically initializes every time Lumaview is started by moving the microplate
carrier to the Home position. The Home Position consists of the left front corner of the microplate
carrier frame being positioned over the objective and the objective retracted to the lowest Z level.
No image can be obtained at the Home position because it is under the microplate carrier frame.
2.
If this is the first time the LS720 has been connected to the current computer, it is first necessary
to do a Calibration.
Once completed, calibration is not needed again unless a different computer is
connected.
a.
To calibrate the LS720, install the provided Calibration Motic EFN 4x
Objective. Place the provided Calibration 1536 well microplate into the
microplate carrier.
b.
Click Utilities in the File Menu bar and in the drop-down menu click
Calibrate to open its dialog box. A communication dialog box will open
asking if a 4x objective has been installed. Click Yes and the communication
dialog box will close.
c.
Note: if your Calibrate menu is offset similar to this: