OMNI-Test Laboratories, Inc.
Control No. C-SFAK-0001, Effective date: 12/2/2015
Page 1 of 1
DIFFERENTIAL PRESSURE GAUGE CALIBRATION DATA SHEET
Instrument to be calibrated:
Pressure Transducer
Maximum Range: 0
– 2” H
2
O
ID Number: OMNI-00371
Calibration Instrument: Digital Manometer
ID Number: OMNI-00633
Date:
11/30/2020
By:
Tony Tong
This form is to be used only in conjunction with Standard Procedure C-SPC.
*Acceptable tolerance is 4%.
The uncertainty of measurement is ±0.4
” WC. This is based on the reference standard having a TAR (Test
Accuracy Ratio) of at least 4:1.
Technician signature:
Date:
Reviewed by:
Date:
Range of
Calibration Point
(
″WC)
Digital
Manometer Input
(
″WC)
Pressure Gauge
Response
(
″WC)
Difference
| Input
–
Response |
% Error of
Full Span
*
0-20% Max. Range
0.0
– 0.4
0.088
0.103
0.015
0.75
20-40% Max. Range
0.4
– 0.8
0.476
0.490
0.014
0.70
40-60% Max. Range
0.8
– 1.2
0.971
0.984
0.013
0.65
60-80% Max. Range
1.2
– 1.6
1.427
1.435
0.008
0.40
80-100% Max. Range
1.6
– 2.0
1.700
1.705
0.005
0.25
11/30/2020
89 of 161
11/30/2020
Содержание BakeHeart
Страница 34: ...34 of 161...
Страница 35: ...35 of 161...
Страница 42: ...42 of 161...
Страница 43: ...43 of 161...
Страница 52: ...52 of 161...
Страница 53: ...53 of 161...
Страница 60: ...60 of 161...
Страница 61: ...61 of 161...
Страница 70: ...70 of 161...
Страница 71: ...71 of 161...
Страница 75: ...75 of 161...
Страница 76: ...76 of 161...
Страница 77: ...77 of 161...
Страница 78: ...78 of 161...
Страница 79: ...79 of 161...
Страница 80: ...80 of 161...
Страница 81: ...81 of 161...
Страница 82: ...82 of 161...
Страница 83: ...83 of 161...
Страница 85: ...85 of 161...
Страница 86: ...86 of 161...
Страница 87: ...87 of 161...
Страница 94: ...94 of 161...
Страница 96: ...96 of 161...
Страница 97: ...97 of 161...
Страница 113: ......
Страница 114: ...Bakeheart and Warmheart Wood Fired Cook Stove Owner s Manual OMNI Test Laboratories Inc 0224WS009S 0224WS009E...
Страница 159: ......