ESS HVP-2100 Скачать руководство пользователя страница 10

 

 

10 

4.       AWB:  

   Press to adjust the White Balance automatically. 

 5.       PARAMETER ADJUSTING/SELECT

(



)

Select the parameter, including brightness, size of the window, contrast, language, 
and color. 
 

6.       Main POWER 

On: [

], OFF: [

], the indicator illuminates when power on. 

7.       SELECT/Enter: 

Select the parameter, including brightness, size of the window, 

8.       PUMP: 

   Press to start or stop air feed 

9.       LAMP:  

Press to illuminate or turn off the lamp. 

10.      SD: 

Slot for standard SD Memory Card for image and video capture save. 

 11.      NEXT: 

           Cycles through images/videos on SD card (after pressing "VIEW"). 

       12.      SWITCH/PLAY: 

           Switch between image capture and video capture functions; starts & stops video  

 

       playback. 

13.      VIEW: 

Press to view images/videos saved on SD card. 

14.      CAPTURE: 
          Capture images/start & stop video recording (same as capture button on 
           endoscopes and optional rigid camera head.) 
15.      SIGNAL SOCKET: 

CCD signal input, connect to endoscope. 

16.     LCD MONITOR:  

Display image 

       17.     LCD MONITOR CONTROLS/ADJUSTMENTS: 
                  Controls to make adjustments to the LCD monitor itself. 
 

       

NOTE:

 

Monitor is pre-set by the factory to operate optimally with the  

        video  endoscopes and optional endoscopy camera 

– DO NOT MAKE  

        CHANGES WITH THESE CONTROLS!  Power to the monitor will be cut  
        when the Main POWER is turned off and/or monitor cable in the back is  
        removed.

 

 

 
 
 
 
 
 
 
 
 
 
 

Содержание HVP-2100

Страница 1: ...ESS HVP 2100 Veterinary All In One Video System INSTRUCTION MANUAL Endoscopy Support Services Inc 3 Fallsview Lane Brewster NY 10509 P 845 277 1700 F 845 277 7300 REV HVPM 032017A...

Страница 2: ...ct order after reading carefully and comprehending the cautions Warning Improper operation is very dangerous Patient operator or person s close to the instrument may be injured Operate this system aft...

Страница 3: ...lug from the socket immediately and contact ESS Inc or your ESS Inc Representative 9 WARNING Never close monitor with anything on top of the HVP 2100 video system this could cause damage to the LCD di...

Страница 4: ...efore Installation 13 3 2 Endoscope Connection 13 3 3 Monitor Connection 14 4 OPERATION INSTRUCTION 14 4 1 Inspection at startup 14 4 2 Parameter Adjustment 15 4 3 Image Video Capture 16 5 MAINTENANCE...

Страница 5: ...l is for the HVP 2100 and HVP 2100L only ESS Inc is not liable for any damage or accident resulting from improper use of the instrument or operation that does not follow the instruction manual The aft...

Страница 6: ...ion caution items about body injury caution items about equipment damage consecutive numbers indicating successive operation note cue or supplement SYMBOLS ON HVP 2100 UNIT Symbol Explanation BF TYPE...

Страница 7: ...magnetic radiation can interfere with the image display 1 2 Content of Warning in Each Chapter Chapter 3 Instrument Installation 3 1 Preparation before Installation Use other power supply voltage than...

Страница 8: ...ervice technician 5 4 Lamp Replacement Operating the equipment before lamp replacement may cause overheating of the lamp Do not replace the lamp instantly to avoid damage to hands Do not touch lamp wh...

Страница 9: ...ideo Cable 1 6 BNC Video Cable 1 7 Fuse T2AL250V 2 2 2 Front Panel Fig 2 1 HVP 2100 Front Panel 1 OUTPUT SOCKET Connect to endoscopy light guide connector 2 BRIGHTNESS Rotate clockwise the level incre...

Страница 10: ...ssing VIEW 12 SWITCH PLAY Switch between image capture and video capture functions starts stops video playback 13 VIEW Press to view images videos saved on SD card 14 CAPTURE Capture images start stop...

Страница 11: ...o ground casing against electrostatic discharge 3 Air intake nozzle for air pump keep clear 4 Monitor video input for flip top LCD monitor 5 RS 232 Serial Port for setup controller 6 Monitor video out...

Страница 12: ...2 3 HVP 2100 Medical Video Center Side Panel 1 Lamp Access Plate open to remove install light source lamp 2 Fix bolt holds lamp access plate closed when in operation 3 Pothook used to hold the water b...

Страница 13: ...und Prevent the air vents from being obstructed to ensure necessary ventilation Insert the power cord into power supply cord receptacle at the back panel 3 2 Endoscope Connection Please refer to endos...

Страница 14: ...other potentially infectious material Warning Do not look directly to the light while the lamp is working The Intense light from endoscope s distal end and light output of the light source may damage...

Страница 15: ...nel under the supervision of a physician and must have received sufficient training in clinical endoscopic technique This manual therefore does not explain or discuss clinical endoscopic procedures It...

Страница 16: ...display language X English Deutsch Fran ais Alter the language COLOR HUE color adjust X from 19 to 19 Alter the level of color are used for adjusting the value of the selected parameter If the paramet...

Страница 17: ...mode press CAPTURE to take videos press again to stop Maximum video capture time is 10 minutes after which without pressing the CAPTURE button the system will stop taking video and store the video mor...

Страница 18: ...se damage to the unit and operational failure Liquids flowing into the equipment may cause fault Do not splash water disinfectant etc onto the equipment and do not force fluids disinfectant with a soa...

Страница 19: ...e equipment by unqualified personnel to avoid personal injury or damage to the medical video center Service or repair should be conducted by a qualified manufacturer service technician Caution Store u...

Страница 20: ...unplug the power cord from the receptacle in back Replace fuses as instructed in following figures Fuse rating is T2AL250V a Remove fuse holder by gently prying it out with a small flathead screwdrive...

Страница 21: ...coupler on rigid endoscopes or flexible fiberoptic endoscopes Camera head endocoupler is out of focus or there is fluids debris on lenses Make sure lenses on both ends of the endocoupler are clean Rot...

Страница 22: ...om 19 to 19 brightness X from 0 to 9 contrast X SOFT MEDIUM HIGH Image frozen 4 images One 1 BNC component video output terminal One 1 Y C video output terminal TV System PAL 625 lines 50Hz Light Sour...

Страница 23: ...cts repaired or modified by a non ESS authorized facility 4 Cautions about repairing For best results ship the video processor to be repaired in the original box and shipping foam If those are not ava...

Страница 24: ...24 NOTES...

Страница 25: ...25 NOTES...

Страница 26: ...26...

Страница 27: ...27...

Страница 28: ...ontinental U S only 1 800 386 3636 Phone 1 845 277 1700 Service Dept ext 214 ESS Inc Main Office Fax 1 845 277 7300 Service Department repairs Fax 1 845 276 4200 ESS website www endoscopy com Service...

Отзывы: