Class II, Type B2 Biological Safety Cabinets 

Chapter 3 – Basic Cabinet Operation 

 

3.1 Sash Window Operation  

3.1.1 Sash Window State 

 

 

 

3.1.2 Operating Motorized Sash Window (Only for BSC with motorized sash window)  

The motorized sash uses a “push and hold” mechanism, so if you remove your finger from the button the sash 
will stop immediately – this is a safety feature to control the closure and prevent anything getting trapped in 
the aperture as the sash descends.  
 

Lower Sash from Fully Open Position 

 

When the sash is fully open, pressing the down button and holding it will cause the sash to move to the Safe 
Height setting and stop. If the fluorescent lights are on as the sash descends, they will stay on as long as the 
sash stops in the Safe Position. If you release the button before the sash has reached Safe Position the lights 
will switch off automatically.  
 

Lower Sash from Safe Height Position 

 

When the sash is at safe operating height pressing the down button and holding it will cause the sash to move 
down to the fully closed position and stop. If the fluorescent lights are on as the sash descends, they will 
switch off automatically as soon as the sash reaches fully closed. If you release the button before the sash has 
reached the fully closed position the lights will switch off automatically 
 

Raise Sash from Fully Closed Position  

When the sash is fully closed, pressing the up button and holding it will prompt the user to input the password 
to turn on the fan. If the password is correct, if it was on fan will turn on and the sash will move up to the Safe 
Height setting and stop.  
 

Raise Sash from Safe Height Position  

When the sash is safe operation position, pressing the up button and holding it will cause the sash to move up 
to the fully open position and stop. If the fluorescent lights are on as the sash rises, they will stay on as long as 
the sash is allowed to fully open. Stopping the sash midway will cause the lights to switch off automatically.

Sash is fully open 

 



Blower can be active



Fluorescent lights can be used



Unsafe working condition

 

Sash is in safe position 

 



Blower can be activated



Fluorescent lights can be used



Safe working condition

 

Sash is fully closed 

 



Blower can’t be activated



Fluorescent lights can’t be used

 

Содержание Labculture LB2 Series

Страница 1: ...Hatboro PA 19040 USA Toll Free USA and Canada 1 877 479 3726 Tel 215 441 9661 Fax 215 441 9660 us escoglobal com usa escoglobal com Esco Micro Pte Ltd 21 Changi South Street 1 Singapore 486 777 Tel 65...

Страница 2: ...change Sentinel and Labculture are registered trademarks of Esco Material in this manual is provided for informational purposes only The contents and the product described in this manual including any...

Страница 3: ...2 2 1 Settings 5 2 2 2 Setting Mode 6 2 2 3 Field Calibration 6 2 2 4 Admin Settings 8 2 3 Stopwatch and Experiment Timer 8 2 4 Alarm and Warnings 9 2 5 Diagnostic Mode 11 Chapter 3 Basic Cabinet Oper...

Страница 4: ...ii...

Страница 5: ...o the product Consumables such as filters HEPA ULPA carbon pre filters and fluorescent UV bulbs Esco is not liable for any damage incurred on the objects used on or stored in Esco equipment If the obj...

Страница 6: ...years for Ascent Max Cleanroom Equipment 1 year limited Laboratory Ovens and Incubators 1 year limited CO2 Incubators 2 years limited Containment Pharma Products 2 years limited Ultralow Temperature...

Страница 7: ...ant safety related points will be marked with the symbol If the equipment is used in a manner not specified by this manual the protection provided by this equipment may be impaired 3 Limitation of Lia...

Страница 8: ...vi...

Страница 9: ...ctive and its amending directives 92 31 EEC The Electromagnetic Compatibility Directive and its amending directives has been designed to comply with the requirement of the following Harmonized Standar...

Страница 10: ...viii...

Страница 11: ...1 Class II Type B2 Biological Safety Cabinets Chapter 1 Product Information 1 1 Quick View 1 Esco Sentinel Control System 2 Sash Window 3 Interior Work Area 4 Armrest 1 2 3 4...

Страница 12: ...2...

Страница 13: ...ase corresponding value inside one of the menu options o Move the sash window upward and downward for motorized sash BSC o Accessing the stopwatch and experiment timer function for non motorized sash...

Страница 14: ...ly available when MOTORIZED WINDOW is chosen 2 2 1 Settings Users may use the settings menu function to customize the operation of the BSC to meet specific application requirements The settings menu c...

Страница 15: ...TIMER 2 2 1 7 Experiment Timer This menu enable the personel to use experiment timer function in a BSC with motorized window See section 2 3 for more information on experiment timer function and secti...

Страница 16: ...the settings menu to their factory settings 2 2 4 1 New ADMIN PIN default 0009 ADMIN PIN restricts access to some of the more delicate menu functions namely admin and field calibration which should on...

Страница 17: ...The counter value can be checked while in the diagnostic mode Please reset the UV lamp hour meter after each UV lamp replacement MENU ADMIN RESET UV H M 2 2 4 6 Reset Default User can reset the defaul...

Страница 18: ...he timer in the experiment timer function is counting down and shown using the HH MM SS format Operator can use the SETTINGS EXPERIMENT TIMER menu refer to section 2 2 1 4 to set the experiment timer...

Страница 19: ...ite on MUTE TIMER Shows how long the airfail alarm is muted AF MONITOR Airflow monitor condition AF OUT TEMP Airflow out of range temperature condition ADC DMIN ADC value for minimum downflow ADC DNOM...

Страница 20: ...10...

Страница 21: ...y closed position and stop If the fluorescent lights are on as the sash descends they will switch off automatically as soon as the sash reaches fully closed If you release the button before the sash h...

Страница 22: ...the fan by pressing the FAN button Input the Fan PIN if asked if PIN 0000 This will start the warm up procedure default 3 minutes All buttons are diabled during warm up period 3 The BSC is ready for w...

Страница 23: ...be operating the BSC in sitting rather than standing posture There are some obvious advantages of the sitting posture The physiological energy cost and fatigue involved in sitting are relatively less...

Страница 24: ...ed position and the interlock is working properly before activating the UV lamp Avoid direct contact with skin and eyes as UV light is classified as a probable human carcinogen The UV lamp should be a...

Страница 25: ...al Corrosive Flammable Skin Irritant Eye Irritant Respiratory Irritant Toxic Use in BSCs Routine Surface Decon Biohazardous Spill 1 protected from light and air effective results may vary blank not ef...

Страница 26: ...16...

Страница 27: ...water afterward Clean the sash window using an appropriate disinfectant agent and glass cleaner afterward Use a damp cloth to clean the exterior surface of the BSC particularly on the front and top in...

Страница 28: ...e sash by continuously until the sash is in the lowest position and create a 10 cm opening between the sash and the diffuser 3 Open the front panel 4 Slide your hand through the opening and clean the...

Страница 29: ...NANCE MODE this mode allows you to lower the sash further than usual 3 Lower the sash by pressing the DOWN button continuously until the sash is in the lowest position and create a 10 cm opening betwe...

Страница 30: ...20...

Страница 31: ...APPENDIX...

Страница 32: ......

Страница 33: ...peration problems encountered etc 2 Any decontamination procedure performed by either the user or the technician should be recorded down as well 3 Please also record any major maintenance procedure pe...

Страница 34: ......

Отзывы: