3.1.3 Using Sash Window  

 

The sash window should be fully closed when the cabinet is not in use. This helps keep the work zone 
interior clean.  

 

The sash window should always be in the normal operating height at all times when the cabinet is in 
use. Even if the cabinet is left unattended, but the blower is on, the sash window should never be 
moved from the normal operating height, unless during loading or unloading of materials/apparatus 
into the cabinet.  

 

The alarm will be activated whenever the sash window is moved from the normal operating height.  

 

Whenever the sash window is moved to the correct height from a higher or lower position, the light 
will automatically be turned on as a signal to the user.  

 

The sash window may be opened to its maximum position for the purpose of loading/unloading of 
materials/apparatus into the cabinet. When the sash window is fully opened, the alarm sound may be 
muted by pressing MUTE button but, will be automatically sounded again after 5 minutes to remind 
the user that it is not safe to work in the cabinet and the light will be turned on to facilitate cleaning.

 

3.2 Starting and Shutting Down the BSC 

3.2.1 Turning on the BSC 

1.

 

Raise the sash to the indicated normal operational height (READY state). The lamp will turned on 
when this height is reached.  

Note:

 When Quickstart mode is selected, fan will turn on as well, without pressing the fan button. 

2.

 

Turn on the fan by pressing the FAN button. Input the Fan PIN if asked (if PIN ≠ 0000). This will start 
the warm up procedure (default: 3 minutes). All buttons are diabled during warm up period.  

3.

 

The BSC is ready for work. 

3.2.2 Turning off the BSC 

1.

 

Turn off the fan by pressing the FAN button. Input the Fan PIN if asked (if PIN ≠ 0000). This will start 
the post purge procedure (default: 0 minute). All buttons are diabled during post purge period. 

2.

 

Lower the sash to the fully closed position (the display will show UV MODE).  The sash can be lowered 
immediately after turning off the fan as it will not interrupt the post purge procedure.  

Note:

 When Quickstart mode is selected, fan will turn off without pressing the fan button.

 

3.

 

Turn on the UV lamp (when present) to decontaminate the work area by pressing the UV button. 
Leave the UV lamp on to make sure the decontamination is done effectively . The UV lamp can 
only be turned on after the post purge procedure is finished. 

3.3 Working in the BSC 

 

Surface-decontaminate the work area (work surface, back and side wall, UV lamp, electrical outlet, 
service fixture and the inner surface of the sash window) before and after using the BSC.  

 

Allow the BSC to purge any contaminant by allowing the blower to operate at least 3 minutes before 
and after using the BSC. 

 

Surface-decontaminate any item (including your arms) or apparatus with appropriate disinfectant 
before entering or exiting the work area. 

 

Place the waste container (containment bag, pipette discard pans, etc.) inside the BSC work area.  

 

Place all items and apparatus in the safe working area. 

Safe Working Area for Single Tray BSC 

 

Содержание Labculture LB2 Series

Страница 1: ...Hatboro PA 19040 USA Toll Free USA and Canada 1 877 479 3726 Tel 215 441 9661 Fax 215 441 9660 us escoglobal com usa escoglobal com Esco Micro Pte Ltd 21 Changi South Street 1 Singapore 486 777 Tel 65...

Страница 2: ...change Sentinel and Labculture are registered trademarks of Esco Material in this manual is provided for informational purposes only The contents and the product described in this manual including any...

Страница 3: ...2 2 1 Settings 5 2 2 2 Setting Mode 6 2 2 3 Field Calibration 6 2 2 4 Admin Settings 8 2 3 Stopwatch and Experiment Timer 8 2 4 Alarm and Warnings 9 2 5 Diagnostic Mode 11 Chapter 3 Basic Cabinet Oper...

Страница 4: ...ii...

Страница 5: ...o the product Consumables such as filters HEPA ULPA carbon pre filters and fluorescent UV bulbs Esco is not liable for any damage incurred on the objects used on or stored in Esco equipment If the obj...

Страница 6: ...years for Ascent Max Cleanroom Equipment 1 year limited Laboratory Ovens and Incubators 1 year limited CO2 Incubators 2 years limited Containment Pharma Products 2 years limited Ultralow Temperature...

Страница 7: ...ant safety related points will be marked with the symbol If the equipment is used in a manner not specified by this manual the protection provided by this equipment may be impaired 3 Limitation of Lia...

Страница 8: ...vi...

Страница 9: ...ctive and its amending directives 92 31 EEC The Electromagnetic Compatibility Directive and its amending directives has been designed to comply with the requirement of the following Harmonized Standar...

Страница 10: ...viii...

Страница 11: ...1 Class II Type B2 Biological Safety Cabinets Chapter 1 Product Information 1 1 Quick View 1 Esco Sentinel Control System 2 Sash Window 3 Interior Work Area 4 Armrest 1 2 3 4...

Страница 12: ...2...

Страница 13: ...ase corresponding value inside one of the menu options o Move the sash window upward and downward for motorized sash BSC o Accessing the stopwatch and experiment timer function for non motorized sash...

Страница 14: ...ly available when MOTORIZED WINDOW is chosen 2 2 1 Settings Users may use the settings menu function to customize the operation of the BSC to meet specific application requirements The settings menu c...

Страница 15: ...TIMER 2 2 1 7 Experiment Timer This menu enable the personel to use experiment timer function in a BSC with motorized window See section 2 3 for more information on experiment timer function and secti...

Страница 16: ...the settings menu to their factory settings 2 2 4 1 New ADMIN PIN default 0009 ADMIN PIN restricts access to some of the more delicate menu functions namely admin and field calibration which should on...

Страница 17: ...The counter value can be checked while in the diagnostic mode Please reset the UV lamp hour meter after each UV lamp replacement MENU ADMIN RESET UV H M 2 2 4 6 Reset Default User can reset the defaul...

Страница 18: ...he timer in the experiment timer function is counting down and shown using the HH MM SS format Operator can use the SETTINGS EXPERIMENT TIMER menu refer to section 2 2 1 4 to set the experiment timer...

Страница 19: ...ite on MUTE TIMER Shows how long the airfail alarm is muted AF MONITOR Airflow monitor condition AF OUT TEMP Airflow out of range temperature condition ADC DMIN ADC value for minimum downflow ADC DNOM...

Страница 20: ...10...

Страница 21: ...y closed position and stop If the fluorescent lights are on as the sash descends they will switch off automatically as soon as the sash reaches fully closed If you release the button before the sash h...

Страница 22: ...the fan by pressing the FAN button Input the Fan PIN if asked if PIN 0000 This will start the warm up procedure default 3 minutes All buttons are diabled during warm up period 3 The BSC is ready for w...

Страница 23: ...be operating the BSC in sitting rather than standing posture There are some obvious advantages of the sitting posture The physiological energy cost and fatigue involved in sitting are relatively less...

Страница 24: ...ed position and the interlock is working properly before activating the UV lamp Avoid direct contact with skin and eyes as UV light is classified as a probable human carcinogen The UV lamp should be a...

Страница 25: ...al Corrosive Flammable Skin Irritant Eye Irritant Respiratory Irritant Toxic Use in BSCs Routine Surface Decon Biohazardous Spill 1 protected from light and air effective results may vary blank not ef...

Страница 26: ...16...

Страница 27: ...water afterward Clean the sash window using an appropriate disinfectant agent and glass cleaner afterward Use a damp cloth to clean the exterior surface of the BSC particularly on the front and top in...

Страница 28: ...e sash by continuously until the sash is in the lowest position and create a 10 cm opening between the sash and the diffuser 3 Open the front panel 4 Slide your hand through the opening and clean the...

Страница 29: ...NANCE MODE this mode allows you to lower the sash further than usual 3 Lower the sash by pressing the DOWN button continuously until the sash is in the lowest position and create a 10 cm opening betwe...

Страница 30: ...20...

Страница 31: ...APPENDIX...

Страница 32: ......

Страница 33: ...peration problems encountered etc 2 Any decontamination procedure performed by either the user or the technician should be recorded down as well 3 Please also record any major maintenance procedure pe...

Страница 34: ......

Отзывы: