19
Class II, Type B2 Biological Safety Cabinets
Sash cleaning procedure for motorized sash BSC
4.2 Maintenance/Service Log
It is good practice (and in some cases regulatory requirement) to maintain a log of all maintenance work
carried out on your cabinet.
1.
Remove armrest
2.
Press MENU button, select SET MODE
then choose MAINTENANCE MODE - this
mode allows you to lower the sash
further than usual
3.
Lower the sash by pressing the DOWN
button continuously until the sash is in
the lowest position and create a 10 cm
opening between the sash and the
diffuser
4.
Open the front panel.
5.
Slide your hand through the opening and
clean the inner side of the sash using
cloth and IPA 70%
6.
After cleaning, push the UP button
continuously until the sash is in the sash
ready position
7.
Close the front panel.
8.
Reattach armrest
9.
Set the cabinet back to NORMAL MODE