2.2.4.10 Set Sash Cycle (Only for Motorized BSC) 

To reset the sash cycles count to zero or to input the sash cycle count manually. The maximum of value of sash 
cycles is 16000. The cycles will raise every sash move to up and down. 
 
Warning message will be showed after the cycle value reach 15000. 
1st warning: “Replace Sash Motor” – after sash reached 15000 cycles. 
2nd warning: “Stop Using Sash” - after sash reached 15500 cycles. 
3rd warning: “Sash Motor Locked” (Sash motor cannot operate anymore) - after sash reached 16000 cycles. 

MENU

ADMIN 

SET SASH CYCLE

RESET SASH CYCLE

SET SASH CYCLE

 

 

2.3 Stopwatch and Experiment Timer (only for non-motorized sash BSC) 

 

The stopwatch function can be started by pressing the UP button while the sash is in the safe/ready 
position. Pressing UP button again while the stopwatch function will stop and resume the timer. 
Pressing DOWN button will leave the stopwatch function and reset the timer. The timer in the 
stopwatch function is counting up and shown using the HH:MM:SS format. 

 

The experiment timer can be started by pressing the DOWN button while the sash is in the safe/ready 
position. Pressing DOWN button while the experiment timer function is working will stop and resume 
the timer. Pressing the UP button will leave the experiment timer function and reset the timer. The 
timer in the experiment timer function is counting down and shown using the HH:MM:SS format. 
Operator can use the SETTINGS | EXPERIMENT TIMER menu (refer to section 2.2.1.4) to set the 
experiment timer. 

2.4 Alarms and Warnings 

BSC uses alarms to indicate that the condition inside the BSC is not safe for the operator, so check the LCD 
display to understand the cause of these alarms. The most common alarm is the SASH ALARM that indicates 
that the sash is neither at the normal operating height nor at fully closed position (UV mode) – this condition 
can easily be corrected by putting the sash at the appropriate operation position.  
 
Another warning that should be acted upon is AIR FAIL! which indicates that there is airflow failure. The 
operator should check if there is any obstruction to the airflow, and correct it if possible. However, if the 
problem continues, the operator should stop working as the BSC’s protection may have been compromised. 
Call service or Esco’s local distributor. 
 
Other alarms that indicate a failure or an error in the BSC system: 

 

AIRFAIL:NO! will be displayed if the blower is turned off while there is an airflow failure. 

 

SASH:ERROR POSITION will be displayed if the microprocessor (controller) detects more than one 
micro-switch activated at the same time, which is impossible, as the sash can only be at one position 
at one time. This indicates a failure in the sash detection system. 

 

SENSOR UNCALIBRATED will be displayed if the airflow velocity sensor is not yet calibrated. 

 

 

Содержание Labculture LB2 Series

Страница 1: ...Hatboro PA 19040 USA Toll Free USA and Canada 1 877 479 3726 Tel 215 441 9661 Fax 215 441 9660 us escoglobal com usa escoglobal com Esco Micro Pte Ltd 21 Changi South Street 1 Singapore 486 777 Tel 65...

Страница 2: ...change Sentinel and Labculture are registered trademarks of Esco Material in this manual is provided for informational purposes only The contents and the product described in this manual including any...

Страница 3: ...2 2 1 Settings 5 2 2 2 Setting Mode 6 2 2 3 Field Calibration 6 2 2 4 Admin Settings 8 2 3 Stopwatch and Experiment Timer 8 2 4 Alarm and Warnings 9 2 5 Diagnostic Mode 11 Chapter 3 Basic Cabinet Oper...

Страница 4: ...ii...

Страница 5: ...o the product Consumables such as filters HEPA ULPA carbon pre filters and fluorescent UV bulbs Esco is not liable for any damage incurred on the objects used on or stored in Esco equipment If the obj...

Страница 6: ...years for Ascent Max Cleanroom Equipment 1 year limited Laboratory Ovens and Incubators 1 year limited CO2 Incubators 2 years limited Containment Pharma Products 2 years limited Ultralow Temperature...

Страница 7: ...ant safety related points will be marked with the symbol If the equipment is used in a manner not specified by this manual the protection provided by this equipment may be impaired 3 Limitation of Lia...

Страница 8: ...vi...

Страница 9: ...ctive and its amending directives 92 31 EEC The Electromagnetic Compatibility Directive and its amending directives has been designed to comply with the requirement of the following Harmonized Standar...

Страница 10: ...viii...

Страница 11: ...1 Class II Type B2 Biological Safety Cabinets Chapter 1 Product Information 1 1 Quick View 1 Esco Sentinel Control System 2 Sash Window 3 Interior Work Area 4 Armrest 1 2 3 4...

Страница 12: ...2...

Страница 13: ...ase corresponding value inside one of the menu options o Move the sash window upward and downward for motorized sash BSC o Accessing the stopwatch and experiment timer function for non motorized sash...

Страница 14: ...ly available when MOTORIZED WINDOW is chosen 2 2 1 Settings Users may use the settings menu function to customize the operation of the BSC to meet specific application requirements The settings menu c...

Страница 15: ...TIMER 2 2 1 7 Experiment Timer This menu enable the personel to use experiment timer function in a BSC with motorized window See section 2 3 for more information on experiment timer function and secti...

Страница 16: ...the settings menu to their factory settings 2 2 4 1 New ADMIN PIN default 0009 ADMIN PIN restricts access to some of the more delicate menu functions namely admin and field calibration which should on...

Страница 17: ...The counter value can be checked while in the diagnostic mode Please reset the UV lamp hour meter after each UV lamp replacement MENU ADMIN RESET UV H M 2 2 4 6 Reset Default User can reset the defaul...

Страница 18: ...he timer in the experiment timer function is counting down and shown using the HH MM SS format Operator can use the SETTINGS EXPERIMENT TIMER menu refer to section 2 2 1 4 to set the experiment timer...

Страница 19: ...ite on MUTE TIMER Shows how long the airfail alarm is muted AF MONITOR Airflow monitor condition AF OUT TEMP Airflow out of range temperature condition ADC DMIN ADC value for minimum downflow ADC DNOM...

Страница 20: ...10...

Страница 21: ...y closed position and stop If the fluorescent lights are on as the sash descends they will switch off automatically as soon as the sash reaches fully closed If you release the button before the sash h...

Страница 22: ...the fan by pressing the FAN button Input the Fan PIN if asked if PIN 0000 This will start the warm up procedure default 3 minutes All buttons are diabled during warm up period 3 The BSC is ready for w...

Страница 23: ...be operating the BSC in sitting rather than standing posture There are some obvious advantages of the sitting posture The physiological energy cost and fatigue involved in sitting are relatively less...

Страница 24: ...ed position and the interlock is working properly before activating the UV lamp Avoid direct contact with skin and eyes as UV light is classified as a probable human carcinogen The UV lamp should be a...

Страница 25: ...al Corrosive Flammable Skin Irritant Eye Irritant Respiratory Irritant Toxic Use in BSCs Routine Surface Decon Biohazardous Spill 1 protected from light and air effective results may vary blank not ef...

Страница 26: ...16...

Страница 27: ...water afterward Clean the sash window using an appropriate disinfectant agent and glass cleaner afterward Use a damp cloth to clean the exterior surface of the BSC particularly on the front and top in...

Страница 28: ...e sash by continuously until the sash is in the lowest position and create a 10 cm opening between the sash and the diffuser 3 Open the front panel 4 Slide your hand through the opening and clean the...

Страница 29: ...NANCE MODE this mode allows you to lower the sash further than usual 3 Lower the sash by pressing the DOWN button continuously until the sash is in the lowest position and create a 10 cm opening betwe...

Страница 30: ...20...

Страница 31: ...APPENDIX...

Страница 32: ......

Страница 33: ...peration problems encountered etc 2 Any decontamination procedure performed by either the user or the technician should be recorded down as well 3 Please also record any major maintenance procedure pe...

Страница 34: ......

Отзывы: