Appendix
0
4
/2019
1019103-00
A
Operating instructions – 87/98
16.1.2
Table 202: Guidelines and manufacturer's
declaration – Electromagnetic interference
The tanning system has been designed for use in the electromagnetic
environment specified below. The device operator must ensure that it is used
in such an environment.
Interference
immunity tests
IEC 60601 -
Test level
Defined
level
Electromagnetic
environment –
Guidelines
Discharging static electricity
(ESD) IEC 61000-4-2
±6 kV contact discharge
±8 kV air discharge
±6 kV contact discharge
±8 kV air discharge
–
Fast transient electrical noise /
bursts IEC 61000-4-4
±2 kV for mains supplies
±2 kV for mains supplies
–
Surges
61000-4-5
±1 kV series mode voltage
±2 kV common mode voltage
±1 kV series mode voltage
±2 kV common mode voltage
–
Magnetic field radiated by
supply frequency (50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
–
Voltage drops,
Short interruptions
and supply voltage
fluctuations
IEC 61000-4-11
<5% U
T
(>95% drop in U
T
)
for 0.5 periods
40% U
T
(60% drop in U
T
)
for 5 periods
70% U
T
(30% drop in U
T
)
for 25 periods
<5% U
T
(>95% drop in U
T
)
for 5 s
<5% U
T
(>95% drop in U
T
)
for 0.5 periods
40% U
T
(60% drop in U
T
)
for 5 periods
70% U
T
(30% drop in U
T
)
for 25 periods
<5% U
T
(>95% drop in U
T
)
for 5 s
–
Note: U
T
is the mains alternating voltage used for the test levels.