What this device is
0
4
/2019
1019103-00
A
Operating instructions – 29/98
JK 126 / 42-4 HT
Power consumption rating:
10,000 W
Rated frequency:
60 Hz
Rated voltage:
230 V 3 phase
230 V 2 phase
Rated fuses:
40 amp 3-pole
60 amp 2-pole
Classification:
Type BF applied part
lying surface (PMMA)
Power supply:
4 x THHN AWG 10
7 x THHN AWG 10
Ingress protection classification:
IP20
Protection against electric shock:
Class I
Max. allowed network impedance:
0.095
Ω
Noise level 1m away from device
(without exhaust air):
68.1 dB(A)
Noise level inside the device:
< 70 dB(A)
Load on the acrylic panel:
max. 330 lb/150 kg
Weight:
approx. 1,190 lb/540 kg
Max. humidity:
< 70%
Ambient temperature:
59-89 °F (15-32 °C)
Max. operating altitude (above sea level):
< 6,500 ft (2,000 m)
NOTE:
Adaptation of the unit is required for operating altitudes
greater than 6,500 ft / 2,000 m above sea level in order to
guarantee trouble-free operation.
Essential performance:
NOTE:
The essential performance of the device is the UV exposure
dose within specified limits, whereas the upper limit must not
be exceeded. No-radiation or zero radition does not result or
contribute to an injury of the user.
Lower limit
exposure dose 0,75 MED
*)
12
Upper limit
exposure dose 4 MED
*)
*) in compliance with 21 CFR 1040.20 and FDA’s timer policy