Maintenance and servicing
801
16-
443
10.
16
Activation with VIO electrosurgical unit (via the ECB)
Testing and measuring equipment
Test setup
•
as with previous testing and also
•
The power cord connects a VIO D or VIO S to the mains power supply.
•
The IES 2 and the VIO are connected to one another via the ECB.
•
Both units are switched on.
•
The two-pedal footswitch with ReMode is connected to the VIO.
Test procedure
Activation via VIO
1. Perform the following settings on the VIO:
– Assign two-pedal footswitch to the bipolar receptacle.
– IES settings for bipolar receptacle: Cut suction: OFF, coag suction
80%, basic suction: OFF, run-on time 1 sec
2. Keep the COAG pedal of the VIO two-pedal footswitch pressed. The
suction must be activated.
3. Release the COAG pedal of the VIO two-pedal footswitch. The suction
must be deactivated.
4. Switch off VIO.
Erbe Art. No.
Description
20322-101
IES 2 one pedal footswitch
20189-303
Dual-pedal footswitch with ReMode
Содержание IES 2
Страница 1: ...10 16 80116 443 SERVICE MANUAL IES 2 1 3 x...
Страница 2: ......
Страница 3: ...SERVICE MANUAL IES 2...
Страница 18: ...4 Installation 18 46 80116 443 10 16...
Страница 19: ...19 46 5 Circuit Descriptions 80116 443 10 16 CHAPTER 5 Circuit Descriptions...
Страница 38: ...8 Maintenance and servicing 38 46 80116 443 10 16...
Страница 40: ...9 Replacing components 40 46 80116 443 10 16...
Страница 45: ...45 46 10 Spare parts 80116 443 10 16 Fig 10 5 40322 009 1 40322 012 1 40322 059 1 only 100 120 V 40322 038 1...
Страница 46: ...10 Spare parts 46 46 80116 443 10 16 Fig 10 6 40322 024 1 only 230 V 40322 025 1 40322 013 1...