8
•
Maintenance and servicing
32
/ 46
801
16-
443
10.
16
Technical safety check – step by step
For simplification the device to be tested is referred to below as the "test
specimen".
Safety information
WARNING!
For safety reasons (personnel protection) the test specimen should
generally be operated by a suitable isolating transformer.
An exceptional case is the tests for grounded conductor resistance,
ground leakage current, and patient leakage current, in which the test
specimen is supplied with current via the safety tester.
ATTENTION!
In the event of a fault occurring in the test specimen or individual com-
ponents during the technical safety check the test steps taken so far no
longer apply.
Remedy the defect and repeat the technical safety check from the
beginning.
IMPORTANT!
It is assumed that the user knows how to operate the test specimen,
the test equipment, the measuring equipment, and auxiliary test equip-
ment. The test instructions only apply in conjunction with the relevant
test steps.
IMPORTANT!
Test equipment, measuring equipment, and auxiliary test equipment
(cables, test boxes, etc.) are listed separately at the beginning of each
test unit. Where Erbe article numbers are specified, only original Erbe
test equipment, measuring equipment, and auxiliary test equipment
may be used.
IMPORTANT!
The test report for the technical safety check can be requested from
Erbe Technical Service Tübingen. For the address see address sheet on
last page.
Содержание IES 2
Страница 1: ...10 16 80116 443 SERVICE MANUAL IES 2 1 3 x...
Страница 2: ......
Страница 3: ...SERVICE MANUAL IES 2...
Страница 18: ...4 Installation 18 46 80116 443 10 16...
Страница 19: ...19 46 5 Circuit Descriptions 80116 443 10 16 CHAPTER 5 Circuit Descriptions...
Страница 38: ...8 Maintenance and servicing 38 46 80116 443 10 16...
Страница 40: ...9 Replacing components 40 46 80116 443 10 16...
Страница 45: ...45 46 10 Spare parts 80116 443 10 16 Fig 10 5 40322 009 1 40322 012 1 40322 059 1 only 100 120 V 40322 038 1...
Страница 46: ...10 Spare parts 46 46 80116 443 10 16 Fig 10 6 40322 024 1 only 230 V 40322 025 1 40322 013 1...