7
•
Remedying malfunctions
30
/ 46
801
16-
443
10.
16
Display y.1
ECB fault
Check IES 2 and VIO electrosurgi-
cal unit software versions for
compatibility.
Display y.2
Excess temperature (> 60 °C)
Switch off the unit and wait until
the temperature in the interior of
the unit has returned to normal.
Display y.3
Insufficient temperature (< 0 °C)
Switch off the unit and wait until
the temperature in the interior of
the unit has returned to normal.
Display y.4
Software error (check total)
Reinstall software.
IMPORTANT!
Software may only
be re-installed and updated by
Erbe Elektromedizin. Send unit for
repair.
Display y.5
Setup parameter error
Restart unit.
Carry out re-initialisation.
Display y.6
Filter data erroneous
Restart unit.
Carry out re-initialisation.
If necessary, replace filter.
Display y.7
Pressure sensor error
Restart unit.
Replace control board.
Error
Possible cause
Action