2
•
Unit description
10
/ 46
801
16-
443
10.
16
Factory setting and Reset
Factory setting
The IES 2 has the following settings when it is delivered by the manufac-
turer:
•
Operation suction: 80 %
•
Basic suction: 10 %
•
Run-on time: 2 s
•
Occlusion / system blocked (sensitivity): 5
•
Occlusion / system blocked (retrigger hold-time): 0 s
•
Display brightness: 10
•
Sound: on (=1)
•
Basic suction time: 30 s
•
Setting mode inhibit: off (=0)
•
Basic setting inhibit: off (=0)
Reset
(reset unit to factory settings)
IMPORTANT!
The unit can only be reset once the "Basic setting inhibit"
service function is off (=0), see page 27.
Fig. 2-1
1. When switching on the unit, keep the power switch and
Operation
suction
button pressed until <S. 0> appears in the display.
2. Press the
START Suction
button within 5 seconds.
The unit is reset to the factory settings if:
– <P. 1> appears in the display,
– the
Basic suction
button illuminates and
– the %-s-display flashes
3. To exit the Setting mode, switch off the unit.
Содержание IES 2
Страница 1: ...10 16 80116 443 SERVICE MANUAL IES 2 1 3 x...
Страница 2: ......
Страница 3: ...SERVICE MANUAL IES 2...
Страница 18: ...4 Installation 18 46 80116 443 10 16...
Страница 19: ...19 46 5 Circuit Descriptions 80116 443 10 16 CHAPTER 5 Circuit Descriptions...
Страница 38: ...8 Maintenance and servicing 38 46 80116 443 10 16...
Страница 40: ...9 Replacing components 40 46 80116 443 10 16...
Страница 45: ...45 46 10 Spare parts 80116 443 10 16 Fig 10 5 40322 009 1 40322 012 1 40322 059 1 only 100 120 V 40322 038 1...
Страница 46: ...10 Spare parts 46 46 80116 443 10 16 Fig 10 6 40322 024 1 only 230 V 40322 025 1 40322 013 1...