Calling up Setting mode
IMPORTANT!
The Setting mode can only be called up once the "Setting
mode inhibit" service function is off (=0), see page 27.
Fig. 6-1
1. When switching on the unit, keep the power switch and
Basic suction
button pressed until <S. 0> appears in the display.
2. Press the
Activation on/off
within 5 seconds.
The Setting mode is called up if:
– <P. 1> appears in the display,
– the
Basic suction
button illuminates and
– the %-s-display flashes
Change settings
1. Select the required setting (P. 1 to P. 6) with the
Higher / Lower but-
tons
.
2. Press the
START Suction
button.
The selected setting can now be changed with the
Higher / Lower
buttons
.
3. To save the changed setting press the
START Suction
button again.
Quitting the setting mode
1. To exit the Setting mode, switch off the unit.
P. 6
Filter
Filter status display
0 – 99 %
99 % (=new)
3 % (=almost
exhausted)
0 %
(=exhausted)
Display
Assembly
Setting
Range
Содержание IES 2
Страница 1: ...10 16 80116 443 SERVICE MANUAL IES 2 1 3 x...
Страница 2: ......
Страница 3: ...SERVICE MANUAL IES 2...
Страница 18: ...4 Installation 18 46 80116 443 10 16...
Страница 19: ...19 46 5 Circuit Descriptions 80116 443 10 16 CHAPTER 5 Circuit Descriptions...
Страница 38: ...8 Maintenance and servicing 38 46 80116 443 10 16...
Страница 40: ...9 Replacing components 40 46 80116 443 10 16...
Страница 45: ...45 46 10 Spare parts 80116 443 10 16 Fig 10 5 40322 009 1 40322 012 1 40322 059 1 only 100 120 V 40322 038 1...
Страница 46: ...10 Spare parts 46 46 80116 443 10 16 Fig 10 6 40322 024 1 only 230 V 40322 025 1 40322 013 1...