160
/ 266
3
CIRCUIT DESCRIPTION
In this way there is another DC voltage, proportional to the spark intensity, which is reduced via the resistor
divider R32, R33 and then fed to the isolation amplifier IC9. The output signal U_FUNKE moves from
there via the control board to control the processor.
The electronic components of this isolation amplifier within the patient circuit require an isolated voltage
supply to function. For this, a DC/DC converter is used which consists of a Colpitts oscillator (circuit
around Transistor T6), the alternating signal of which is transferred from the transformer UE6 to the secondary
side and rectified by diodes D4 and D6. The DC voltages thus resulting are limited to 9.1 volts by the Z-
diodes D3 and D5 respectively.
In this way, voltages of +9.1 volts and –9,1 volts are available to the electronic circuit within the patient
circuit which is identified with I+9 and I–9.
Safety system for the neutral electrode (NESSY)
Problem description: The line for the neutral electrode must be monitored for wire break according to the
standard since there is risk of topical burns for the patient with an incorrect supply line.
In the ICC, correct application of the neutral electrode on the patient is also checked, as well as the uniform
distribution of the HF current to the two electrode surfaces (symmetry).
To monitor the connecting line, the cable is designed as a two-wire connection, the loop resistance of which
is measured. If the resistance value determined in this way is below a specific limit value, one can assume
that the neutral electrode line has no wire break. With conventional single-surface neutral electrodes,
monitoring is performed in this way.
However, such monitoring cannot determine whether the neutral electrode is actually applied to the patient
or whether it is lying around somewhere else.
To also check the correct application to the patient, the contacts of the neutral electrode must be divided,
whereby a control current must flow over the electrode halves and the patient. This control current permits
resistance measurement of the tissue impedance of the patient and thus information as to whether
•
the electrode is not connected or has a wire break (impedance > 120 ohms),
•
is lying directly on the operating table (impedance < 5 ohms),
•
or, most probably, is correctly applied to the patient (impedance between 5 and 120 ohms).
If the high-frequency current is not flowing regularly over both surfaces of a divided neutral electrode, this
can lead to burns due to current concentration under the electrode. It is therefore also necessary to check the
symmetrical current distribution on both halves of the electrode.
The NESSY system, a neutral electrode safety system patented by ERBE, takes care of all the tasks
mentioned.
Principle of the NESSY measurement system
The electric resistance of the neutral electrode is measured via an oscillator, the amplitude of which is
dampened by the resistance from the patient and measured by the system. With a correct circuit design, a
sufficiently precise resistance measurement is possible in this way.
Senso-board
Slot J7
Содержание ICC 200
Страница 1: ...Service manual ERBE ICC 200 ICC 300 H E ICC 350 09 2004...
Страница 2: ......
Страница 6: ...6 266...
Страница 7: ...Chapter1 Test programs and adjustments...
Страница 8: ......
Страница 42: ...1 TEST PROGRAMS AND ADJUSTMENTS 42 266...
Страница 52: ...1 TEST PROGRAMS AND ADJUSTMENTS 52 266...
Страница 53: ...1 TEST PROGRAMS AND ADJUSTMENTS 53 266 Art No 80116 201 09 2004 Adjustment 1...
Страница 56: ...1 TEST PROGRAMS AND ADJUSTMENTS 56 266...
Страница 57: ...1 TEST PROGRAMS AND ADJUSTMENTS 57 266 Art No 80116 201 09 2004 Adjustment 2...
Страница 60: ...1 TEST PROGRAMS AND ADJUSTMENTS 60 266...
Страница 61: ...1 TEST PROGRAMS AND ADJUSTMENTS 61 266 Art No 80116 201 09 2004 Adjustment 3...
Страница 64: ...1 TEST PROGRAMS AND ADJUSTMENTS 64 266...
Страница 65: ...1 TEST PROGRAMS AND ADJUSTMENTS 65 266 Art No 80116 201 09 2004 Adjustment 4...
Страница 68: ...1 TEST PROGRAMS AND ADJUSTMENTS 68 266...
Страница 69: ...1 TEST PROGRAMS AND ADJUSTMENTS 69 266 Art No 80116 201 09 2004 Adjustment 5...
Страница 72: ...1 TEST PROGRAMS AND ADJUSTMENTS 72 266...
Страница 73: ...1 TEST PROGRAMS AND ADJUSTMENTS 73 266 Art No 80116 201 09 2004 Adjustment 6...
Страница 76: ...1 TEST PROGRAMS AND ADJUSTMENTS 76 266...
Страница 77: ...1 TEST PROGRAMS AND ADJUSTMENTS 77 266 Art No 80116 201 09 2004 Adjustment 7...
Страница 79: ...1 TEST PROGRAMS AND ADJUSTMENTS 79 266 Art No 80116 201 09 2004 Adjustment 7 Phase angle cos j...
Страница 80: ...1 TEST PROGRAMS AND ADJUSTMENTS 80 266...
Страница 81: ...1 TEST PROGRAMS AND ADJUSTMENTS 81 266 Art No 80116 201 09 2004 Adjustment 8...
Страница 83: ...1 TEST PROGRAMS AND ADJUSTMENTS 83 266 Art No 80116 201 09 2004 Adjustment 8 Spark monitor...
Страница 84: ...1 TEST PROGRAMS AND ADJUSTMENTS 84 266...
Страница 85: ...1 TEST PROGRAMS AND ADJUSTMENTS 85 266 Art No 80116 201 09 2004 Adjustment 9...
Страница 87: ...1 TEST PROGRAMS AND ADJUSTMENTS 87 266 Art No 80116 201 09 2004 Adjustment 9 NESSY resistance measurement...
Страница 88: ...1 TEST PROGRAMS AND ADJUSTMENTS 88 266...
Страница 89: ...1 TEST PROGRAMS AND ADJUSTMENTS 89 266 Art No 80116 201 09 2004 Adjustment 10...
Страница 92: ...1 TEST PROGRAMS AND ADJUSTMENTS 92 266...
Страница 93: ...1 TEST PROGRAMS AND ADJUSTMENTS 93 266 Art No 80116 201 09 2004 Adjustment 11...
Страница 96: ...1 TEST PROGRAMS AND ADJUSTMENTS 96 266...
Страница 97: ...1 TEST PROGRAMS AND ADJUSTMENTS 97 266 Art No 80116 201 09 2004 Adjustment 12...
Страница 100: ...1 TEST PROGRAMS AND ADJUSTMENTS 100 266...
Страница 101: ...1 TEST PROGRAMS AND ADJUSTMENTS 101 266 Art No 80116 201 09 2004 Adjustment 13...
Страница 103: ...1 TEST PROGRAMS AND ADJUSTMENTS 103 266 Art No 80116 201 09 2004 Adjustment 13 LF leakage current monitor TP1...
Страница 104: ...1 TEST PROGRAMS AND ADJUSTMENTS 104 266...
Страница 110: ...1 TEST PROGRAMS AND ADJUSTMENTS 110 266...
Страница 111: ...1 TEST PROGRAMS AND ADJUSTMENTS 111 266 Art No 80116 201 09 2004 Adjustment of remote control for Neurotest...
Страница 114: ...1 TEST PROGRAMS AND ADJUSTMENTS 114 266...
Страница 115: ...1 TEST PROGRAMS AND ADJUSTMENTS 115 266 Art No 80116 201 09 2004 Adjustment of activation and instrument detection...
Страница 122: ...1 TEST PROGRAMS AND ADJUSTMENTS 122 266...
Страница 127: ...Chapter2 ERROR list...
Страница 128: ......
Страница 136: ...2 ERROR LIST 136 266...
Страница 137: ...Chapter3 Circuit description...
Страница 138: ......
Страница 168: ......
Страница 169: ...Chapter4 Block diagrams...
Страница 170: ......
Страница 174: ......
Страница 175: ...Chapter5 Circuit diagrams...
Страница 176: ......
Страница 177: ...PCBs for ICC 200 300 350...
Страница 181: ...5 CIRCUIT DIAGRAMS 181 266 Art No 80116 201 09 2004 ICC 200 300 350 CPU PCB bare 40128 026...
Страница 188: ...188 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Control board bare 40128 095...
Страница 190: ...190 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 QC Power Stage bare 40128 127...
Страница 192: ...192 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Power Module bare Power Module UL bare 40128 117...
Страница 194: ...194 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 ST Power Stage bare 40128 129...
Страница 198: ...198 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Senso board bare 40128 130...
Страница 200: ...200 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Upper Wiring Module bare 40128 118...
Страница 201: ...PCBs only for ICC 200...
Страница 202: ...202 266 5 CIRCUIT DIAGRAMS...
Страница 208: ...208 266 5 CIRCUIT DIAGRAMS ICC 200 Motherboard bare 40128 025...
Страница 210: ...210 266 5 CIRCUIT DIAGRAMS ICC 200 Mono Output bare 40128 096...
Страница 211: ...PCBs only for ICC 300...
Страница 212: ...212 266 5 CIRCUIT DIAGRAMS...
Страница 218: ...218 266 5 CIRCUIT DIAGRAMS ICC 300 Motherboard bare 40128 092...
Страница 220: ...220 266 5 CIRCUIT DIAGRAMS ICC 300 Relay Board bare 40128 093...
Страница 221: ...PCBs only for ICC 350...
Страница 222: ...222 266 5 CIRCUIT DIAGRAMS...
Страница 229: ...5 CIRCUIT DIAGRAMS 229 266 Art No 80116 201 09 2004 ICC 350 Motherboard bare 40128 092...
Страница 231: ...5 CIRCUIT DIAGRAMS 231 266 Art No 80116 201 09 2004 ICC 350 Relay Board bare 40128 093...
Страница 234: ...234 266 5 CIRCUIT DIAGRAMS ICC 350 Neurotest Neurotest Board 40128 027...
Страница 236: ...236 266 5 CIRCUIT DIAGRAMS ICC 350 Neurotest Motherboard Neurotest bare 40128 028...
Страница 238: ...238 266 5 CIRCUIT DIAGRAMS ICC 350 MIC MIEN DOKU MIC Board bare 40128 064...
Страница 240: ...240 266 5 CIRCUIT DIAGRAMS ICC 350 MIC MIEN DOKU Protection Board bare 40128 141...
Страница 241: ...AppendixA Part Numbers...
Страница 242: ......
Страница 263: ...Appendix B Abbreviations Notes Addresses...
Страница 264: ...264 266 B APPENDIX B...
Страница 266: ...Your notes...