CIRCUIT DESCRIPTION
Art. No.: 801
16-201
0
9
/ 200
4
an inductivity of the coils results which limits the high-frequency leakage current with this inductive
impedance.
Monitoring the high-frequency leakage currents
As intended, the high-frequency current should only be used for cutting or coagulating tissue. To do this, it
must flow to the active electrode in a small area and with a high current density, and flow back again from
the neutral electrode over a large area and with a low current density.
However, if the body of the patient is in contact with another electrically conductive object, the current may
also take an undesirable path and, for example, cause a burn there. These undesirable currents are known as
“leakage currents”. It is therefore important to know whether such a leakage current is flowing and how
large it is.
Such leakage current cannot simply disappear, but must instead always flow back to the generator. The
largest proportion of these high-frequency leakage currents flow back to the generator via the power cable
and the grounded conductor.
Since all power lines and the grounded conductor are directed through the ring core of the coil L1, a
corresponding high frequency voltage is induced if HF leakage current is present in the coil L1, which is
rectified with diode D1, restricted with D6, D7 and smoothed in capacitor C13. For further processing, this
DC voltage is fed to the processor system via the isolation amplifier IC12.
Changeover between flowing and capacitance-grounded output
High-frequency surgical units are generally available either in “capacitance ground of the neutral electrode”
or “floating neutral electrode” connection systems, whereby no components are integrated between the
connection to the neutral electrode and chassis in the latter connection system.
Both connection systems have advantages and disadvantages.
The advantage of capacitance ground is that the feed line from the neutral electrode to the patient is grounded
at a high frequency via the capacitor and the risk of burns is reduced. However, it is a disadvantage that
impermissible low-frequency leakage currents may flow through this capacitor during cardiac operations.
If the neutral electrode is operated as floating, on the other hand, the low-frequency leakage currents can be
kept very easily below the limit value of 10 or 50 mA during cardiac operations, but the risk of burns is
increased in this connection system since the line from the neutral electrode is about half the high frequency
voltage of the generator.
With the ICC high-frequency surgical units, the low-frequency leakage current to the patient is measured.
At the first malfunction, this must not exceed 50 mA. Once this condition is ensured, the unit is permitted
to carry the CF designation (Cardiac Floating). If the possible low-frequency leakage current is under 500
mA in the first case of malfunction, the unit is therefore not suitable for operations on the heart and has the
designation BF (Body Floating).
Initially the ICC is capacitance-grounded via capacitors C30 and C31 and the relay contact REL 2. The
advantages of capacitance ground are fully exploited. The low-frequency leakage current is measured via
the prescribed simulation R25, C28. If for any reason a low-frequency leakage current should appear, it is
then evaluated. If it exceeds the limit of 50 mA, the relay contacts REL 2 are then opened and thus a low-
frequency leakage current can no longer flow. The ICC thus fulfills the CF requirement for cardiac surgery
Motherboard
Содержание ICC 200
Страница 1: ...Service manual ERBE ICC 200 ICC 300 H E ICC 350 09 2004...
Страница 2: ......
Страница 6: ...6 266...
Страница 7: ...Chapter1 Test programs and adjustments...
Страница 8: ......
Страница 42: ...1 TEST PROGRAMS AND ADJUSTMENTS 42 266...
Страница 52: ...1 TEST PROGRAMS AND ADJUSTMENTS 52 266...
Страница 53: ...1 TEST PROGRAMS AND ADJUSTMENTS 53 266 Art No 80116 201 09 2004 Adjustment 1...
Страница 56: ...1 TEST PROGRAMS AND ADJUSTMENTS 56 266...
Страница 57: ...1 TEST PROGRAMS AND ADJUSTMENTS 57 266 Art No 80116 201 09 2004 Adjustment 2...
Страница 60: ...1 TEST PROGRAMS AND ADJUSTMENTS 60 266...
Страница 61: ...1 TEST PROGRAMS AND ADJUSTMENTS 61 266 Art No 80116 201 09 2004 Adjustment 3...
Страница 64: ...1 TEST PROGRAMS AND ADJUSTMENTS 64 266...
Страница 65: ...1 TEST PROGRAMS AND ADJUSTMENTS 65 266 Art No 80116 201 09 2004 Adjustment 4...
Страница 68: ...1 TEST PROGRAMS AND ADJUSTMENTS 68 266...
Страница 69: ...1 TEST PROGRAMS AND ADJUSTMENTS 69 266 Art No 80116 201 09 2004 Adjustment 5...
Страница 72: ...1 TEST PROGRAMS AND ADJUSTMENTS 72 266...
Страница 73: ...1 TEST PROGRAMS AND ADJUSTMENTS 73 266 Art No 80116 201 09 2004 Adjustment 6...
Страница 76: ...1 TEST PROGRAMS AND ADJUSTMENTS 76 266...
Страница 77: ...1 TEST PROGRAMS AND ADJUSTMENTS 77 266 Art No 80116 201 09 2004 Adjustment 7...
Страница 79: ...1 TEST PROGRAMS AND ADJUSTMENTS 79 266 Art No 80116 201 09 2004 Adjustment 7 Phase angle cos j...
Страница 80: ...1 TEST PROGRAMS AND ADJUSTMENTS 80 266...
Страница 81: ...1 TEST PROGRAMS AND ADJUSTMENTS 81 266 Art No 80116 201 09 2004 Adjustment 8...
Страница 83: ...1 TEST PROGRAMS AND ADJUSTMENTS 83 266 Art No 80116 201 09 2004 Adjustment 8 Spark monitor...
Страница 84: ...1 TEST PROGRAMS AND ADJUSTMENTS 84 266...
Страница 85: ...1 TEST PROGRAMS AND ADJUSTMENTS 85 266 Art No 80116 201 09 2004 Adjustment 9...
Страница 87: ...1 TEST PROGRAMS AND ADJUSTMENTS 87 266 Art No 80116 201 09 2004 Adjustment 9 NESSY resistance measurement...
Страница 88: ...1 TEST PROGRAMS AND ADJUSTMENTS 88 266...
Страница 89: ...1 TEST PROGRAMS AND ADJUSTMENTS 89 266 Art No 80116 201 09 2004 Adjustment 10...
Страница 92: ...1 TEST PROGRAMS AND ADJUSTMENTS 92 266...
Страница 93: ...1 TEST PROGRAMS AND ADJUSTMENTS 93 266 Art No 80116 201 09 2004 Adjustment 11...
Страница 96: ...1 TEST PROGRAMS AND ADJUSTMENTS 96 266...
Страница 97: ...1 TEST PROGRAMS AND ADJUSTMENTS 97 266 Art No 80116 201 09 2004 Adjustment 12...
Страница 100: ...1 TEST PROGRAMS AND ADJUSTMENTS 100 266...
Страница 101: ...1 TEST PROGRAMS AND ADJUSTMENTS 101 266 Art No 80116 201 09 2004 Adjustment 13...
Страница 103: ...1 TEST PROGRAMS AND ADJUSTMENTS 103 266 Art No 80116 201 09 2004 Adjustment 13 LF leakage current monitor TP1...
Страница 104: ...1 TEST PROGRAMS AND ADJUSTMENTS 104 266...
Страница 110: ...1 TEST PROGRAMS AND ADJUSTMENTS 110 266...
Страница 111: ...1 TEST PROGRAMS AND ADJUSTMENTS 111 266 Art No 80116 201 09 2004 Adjustment of remote control for Neurotest...
Страница 114: ...1 TEST PROGRAMS AND ADJUSTMENTS 114 266...
Страница 115: ...1 TEST PROGRAMS AND ADJUSTMENTS 115 266 Art No 80116 201 09 2004 Adjustment of activation and instrument detection...
Страница 122: ...1 TEST PROGRAMS AND ADJUSTMENTS 122 266...
Страница 127: ...Chapter2 ERROR list...
Страница 128: ......
Страница 136: ...2 ERROR LIST 136 266...
Страница 137: ...Chapter3 Circuit description...
Страница 138: ......
Страница 168: ......
Страница 169: ...Chapter4 Block diagrams...
Страница 170: ......
Страница 174: ......
Страница 175: ...Chapter5 Circuit diagrams...
Страница 176: ......
Страница 177: ...PCBs for ICC 200 300 350...
Страница 181: ...5 CIRCUIT DIAGRAMS 181 266 Art No 80116 201 09 2004 ICC 200 300 350 CPU PCB bare 40128 026...
Страница 188: ...188 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Control board bare 40128 095...
Страница 190: ...190 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 QC Power Stage bare 40128 127...
Страница 192: ...192 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Power Module bare Power Module UL bare 40128 117...
Страница 194: ...194 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 ST Power Stage bare 40128 129...
Страница 198: ...198 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Senso board bare 40128 130...
Страница 200: ...200 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Upper Wiring Module bare 40128 118...
Страница 201: ...PCBs only for ICC 200...
Страница 202: ...202 266 5 CIRCUIT DIAGRAMS...
Страница 208: ...208 266 5 CIRCUIT DIAGRAMS ICC 200 Motherboard bare 40128 025...
Страница 210: ...210 266 5 CIRCUIT DIAGRAMS ICC 200 Mono Output bare 40128 096...
Страница 211: ...PCBs only for ICC 300...
Страница 212: ...212 266 5 CIRCUIT DIAGRAMS...
Страница 218: ...218 266 5 CIRCUIT DIAGRAMS ICC 300 Motherboard bare 40128 092...
Страница 220: ...220 266 5 CIRCUIT DIAGRAMS ICC 300 Relay Board bare 40128 093...
Страница 221: ...PCBs only for ICC 350...
Страница 222: ...222 266 5 CIRCUIT DIAGRAMS...
Страница 229: ...5 CIRCUIT DIAGRAMS 229 266 Art No 80116 201 09 2004 ICC 350 Motherboard bare 40128 092...
Страница 231: ...5 CIRCUIT DIAGRAMS 231 266 Art No 80116 201 09 2004 ICC 350 Relay Board bare 40128 093...
Страница 234: ...234 266 5 CIRCUIT DIAGRAMS ICC 350 Neurotest Neurotest Board 40128 027...
Страница 236: ...236 266 5 CIRCUIT DIAGRAMS ICC 350 Neurotest Motherboard Neurotest bare 40128 028...
Страница 238: ...238 266 5 CIRCUIT DIAGRAMS ICC 350 MIC MIEN DOKU MIC Board bare 40128 064...
Страница 240: ...240 266 5 CIRCUIT DIAGRAMS ICC 350 MIC MIEN DOKU Protection Board bare 40128 141...
Страница 241: ...AppendixA Part Numbers...
Страница 242: ......
Страница 263: ...Appendix B Abbreviations Notes Addresses...
Страница 264: ...264 266 B APPENDIX B...
Страница 266: ...Your notes...