52
ULTRA-E-F
Notes:
1.
Keep transducer end face and bracket end face at the same plane.
2.
When using screw to fix transducer, we can observe the installation from broadside.
3.
Do not over tighten to prevent couplant extrusion.
Mounting Frame Installation for Dedicated Flowmeter (Flat view)
Mounting Frame Installation for Portable/Handheld Flowmeter (Flat view)
Installation Steps for Z and N Transducer Mounting Methods
When using Z or N method to install transducers, install the two transducers respectively on
the opposite sides of pipeline. Installation steps are as follow.
A, Install metal chain
Refer to step C in steps for V and W method. Due to install the two transducers on the opposite
sides of pipeline, do not use guide rods and graduated scale.
B, Fix bracket to pipeline
Refer to step D in steps for V and W method. Use latch and metal chain to fix bracket on