ULTRA-E-F
Piping configuration
And transducer position
Upstream Dimension
Downstream Dimension
Pipe Diameters (*)
Pipe Diameters (**)
24
5
14
5
10
5
10
5
8
5
24
5
Transducer Spacing
ULTRA-E-F transducers are clamped on the outside of a closed pipe
at a specific distance
from each other
. The transducers can be mounted in V-mode where the sound transverses the
pipe two times, W-mode where the sound transverses the pipe four times, or in Z-mode where
the transducers are mounted on opposite sides of the pipe and the sound crosses the pipe once.
W-mode provides the longest sound path length between the transducers - but the weakest
signal strength. Z-mode provides the strongest signal strength - but has the shortest sound path
length.