15
ULTRA-E-F
4.
Conduit holes. Conduit hubs should be used where cables enter the enclosure. Holes not
used for cable entry should be sealed with plugs.
NOTE: Use NEMA 4 [IP65] rated fittings/plugs to maintain the water tight integrity of the
enclosure. Generally, the left conduit hole (viewed from front) is used for line power; the
center conduit hole for transducer connections and the right hole are utilized for OUTPUT
wiring.
5 If additional holes are required, drill the appropriate size hole in the enclosure’s bottom.
Use extreme care not to run the drill bit into the wiring or circuit cards.
Transmitter Wirings
Figure 3.2 Transmitter Wirings
To access terminal strips for electronic connectors, loosen the two screws in the enclosure door
and open the door. Guide the transducer terminations through the transmitter conduit hole
located in the bottom-center of the enclosure.
The terminals within ULTRA-E-F are a pluggable type - they can be removed wired and then
plugged back in. Connect the appropriate wires to the corresponding screw terminals in the
transmitter.
XDCR UP “+ - GND” is used to connect upstream transducer (Marked with a red sign),