18
ULTRA-E-F
ENTER
Generally, press
key first if operator wants to enter “modify” condition. If the
“modify” is still not possible even after pressing the ENTER key, it means that system is
locked by a password. To “Unlock” it,
Select “Unlock” in Window No. 47 and enter the original password. The keypad will not
respond if the keypad is locked. It only can be unlocked by the entering original password.
Select keypad lock functions in Window No. 48. Please consult factory for password if
necessary.
ULTRA-E-F Window Descriptions
The ULTRA-E-F has a unique feature of windows processing for all
operations. These windows are assigned as follows:
00
~
08 windows for the display of flow rate, velocity, positive total, negative total, net total,
heat flow, date & time, meter run status etc.
11
~
29 windows for initial Parameter Setup: To enter pipe outside diameter, pipe wall
thickness, pipe material type, fluid type, transducer type, etc. For ULTRA-E-F, pipe material
type selection is not necessary.
30
~
38 windows for flow Units Options: to select the flow unit, totalizer unit, measurement
unit, turn totalizers on/off and reset totalizes, etc.
40
~
49
windows for Setup options: Scale factor, network IDN (Window No.46), system
lock (Window No.47) and keypad lock code (Window No.48), etc.
50
~
89 windows for Input and output setup: relay output setup, 4-20mA outputs, flow batch
controller, LCD backlit option, date and time, low/high output frequency, alarm output, date
totalizer, etc.
90
~
94 windows for Diagnoses: Signal strength and signal quality (Window No.90),
TOM/TOS*100 (Window No.91)
, flow sound velocity (Window No.92), total time and delta
time (Window No.93), Reynolds number and factor (Window No.94), etc.
∧
0
~∧
8: Power on/off time, total working hours, on/off times and hardware adjustment,
used by the manufacturer only. For further information, please refer to
Part 4 – Windows
Display Explanations
. If you have any questions, refer to the step-by-step instructions found
in the following section (
3.3.4 Pipe Parameter Entry Shortcuts
).
In fact, users don’t need to set up so many steps, just select necessary parameters to set up in
the menu.
Pipe Parameter Entry Shortcuts
The following parameters should be entered for normal measurement:
1.
Pipe outer diameter
2.
Pipe wall thickness
3.
Pipe material
4.
Liner material parameters (including thickness and sound velocity, if needed)
5.
Fluid type