31
The edited data is transmitted to the Evaluation unit via “Device program”.
Critical data concerning the ModBus communication and the calibration must be confirmed
before the parameters are transmitted to the Evaluation unit:
✔
If, when saving the parameters in the Evaluation unit, the system calibration data is
changed, this action must be confirmed by checking “Overwrite calibration”.
✔
If, when saving the parameters in the Evaluation unit, the system interface parameter are
changed, this must be confirmed by checking the selection “Overwrite baud r./address”.
8.3.5 On-Line representation
By clicking on “On-Line representation” button you will make a new windows appear.
It will display: Sensor RAW value
Measured value
Sensor(s) status
(A file should be created in Data-logger settings first)
In addition, with the PC software,
•
the Evaluation unit parameters of the Evaluation unit can be saved in a file (Save
configuration)
•
the Evaluation unit parameters can be loaded from a file (Load configuration)
•
the Evaluation unit parameters can be printed via the Windows standard printer (Print
configuration)
•
the measured values can be logged in a data logger file (enter the file name and storage
rate, and activate the data logger on the online display)
The software language can be set by right-clicking the “Sprache/Language/Langue” field in the
bottom program line on “Deutsch/English/Français”.
Protection against unauthorised use:
The PC interface does not have a password prompt as it is assumed that only authorised
personnel will have access to the PC and the software. The password to operate the display can
be read and changed in menu
7. System
under
7.7 Password
.