20
4.
Analog output
Setting and calibrating the analog output.
4.1
Lower limit
Input:
0 … 22 mA
Standard setting: 4 mA
4.2
Upper limit
Input:
0 … 22 mA
Standard setting: 20 mA
4.3
Alarm value
Input:
0 … 22 mA
Value to be output at pending alarm
(Standard setting 2 mA)
4.4
Alarm mode
Selection:
Alarm:
Alarm is output
Alarm / Hold output
measurement value drops to 0, or
current measurement value.
Hold output:
Last measurement value
remains pending until fault rectification
at the output signal.
4.5
Analog output 1
Submenu
4.5.1 Calibration 4 mA
Selection:
Key functions can be used to set the
Setting the output current current and equalise it to the receiver side
4.5.2 Calibration 20 mA
Selection:
Key functions can be used to set the
Setting the output current current and equalise it to the receiver side