23
6.
Current input
Note: Only available on MSE 300-FH.
Option for auto-correction by external current signal.
The signal is not electrically isolated.
If the connection is incorrect, the CPU of the Evaluation unit may be destroyed.
An external, galvanic isolation by means of a current disconnector or similar must be
provided.
6.1
Input calib. 4 mA
Selection:
The 4 mA signal must be read in via key
Set input current
functions.
6.2
Input calib. 20 mA
Selection:
The 20 mA signal must be read in via key
Set input current
functions.
6.3 Correction
Selection:
ON:
Activation of the correction.
ON / OFF
OFF:
Deactivation of the correction.
6.4
P1 input
Input:
4 mA … 20 mA
Input the current that
is to be used for the correction.
6.5
P1 factor
Input:
0.01 … 10
Factor for subsequent adjustment of the
actual measurement value.
6.n
Pn input
Input:
4 mA … 20 mA
Option for further entry of current value
and correction factors.
6.n
Pn factor
Input:
0.01 … 10