Page 5 of 14 

EN109-1560750-44

 

4

 

Installation 

Unpacking the device  

  Do not lift the couch by the table top parts or the all-around control. 

The device is heavy, it must be unpacked by at least 2 persons. 

 

Inspection 

  Immediately after unpacking the couch, check whether the package contains all the items 

listed above, and check the couch for any damage. 
In case of damage, contact your local representative. Do NOT use the equipment! 

 

Installation 

  Read this Instructions for Use carefully before using the equipment. 

  Avoid exposure to direct sunlight, UV radiation, rain, excessive dust, moisture, mechanical 

vibrations, and shocks. 

  The Manumed ST couches are not intended for use in so-

called “wet rooms” (hydrotherapy 

rooms). 

  The device may not be used in close proximity (i.e. less than 2 metres) to shortwave or device. 

  The Manumed ST should be placed on a flat surface. 

  Turn the adjustable foot until it touches the ground and the couch is standing stable. 
  For correct assembly and safe use of possible optional accessories one must consult the 

separately provided documents included with the optional accessories. 

 

 
 
 

5

 

Intended use and intended user 

Intended use 

The Manumed ST examination and treatment couches are intended to support a patient in the 
desired position for examination or treatment within the physiotherapy, rehabilitation and adjacent 
disciplines. 
 

Intended user 

The Manumed ST treatment couches are intended to be used, and shall only be used, by or under 
the supervision of professional users in physiotherapy, rehabilitation and adjacent areas. 

 
 
 

Connection to mains supply 

  Mains  supply  connections  must  comply  with  the  national  requirements  regarding  medical 

rooms. 

  Prior to connection of this device to the mains supply, check that the voltage and frequency 

stated on the type plate correspond with the available mains supply. 

Содержание Manumed ST

Страница 1: ...Manumed ST EN109 1560750 44 APRIL 4 2022 Instructions for use...

Страница 2: ...3 2 Symbols 3 3 Package contents 4 4 Installation 5 5 Intended use and intended user 5 6 Precautionary instructions 6 7 Operation 7 8 Safety instructions 9 9 Cleaning maintenance and troubleshooting...

Страница 3: ...o Manumed ST 2 section head section Manumed ST 2 section back section Manumed ST 3 section Manumed ST accessories 2 Symbols Symbol used Description Follow the instructions in the Instructions for Use...

Страница 4: ...d water by mechanical casings and electrical enclosures Double Insulation Type B applied part complying with the specified requirements to provide protection against electric shock particularly regard...

Страница 5: ...d ST should be placed on a flat surface Turn the adjustable foot until it touches the ground and the couch is standing stable For correct assembly and safe use of possible optional accessories one mus...

Страница 6: ...nt to be used by competent personnel in physiotherapy rehabilitation or adjacent areas Avoid any use of the equipment by unauthorized personnel When used by children supervision by a knowledgeable par...

Страница 7: ...upports the full weight of the patient Each section of this Manumed ST is supported by gas springs As well as easing operation they offer full support of the sections However these gas springs are not...

Страница 8: ...ed underneath the framework with your foot the required height of the couch can be set At the head section of the couch the height can be adjusted by turning the handle to the left or to the right Nev...

Страница 9: ...eside the couch is set in flexion position combined with the lowest possible position Underneath the head section between the head section and floor Danger of tilting Do not sit on the head and leg se...

Страница 10: ...clean water Disinfection Clean the couch as described above before proceeding to disinfect the couch Follow the disinfectant s instructions for use Never use a mixture of disinfectants or a disinfecta...

Страница 11: ...service life This device will remain suitable for its intended use as long as it is subjected to yearly maintenance by a qualified service engineer as described in the service manual and the service...

Страница 12: ...EN ISO 10993 5 DIN EN ISO 10993 10 Technical modifications reserved Safety and performance standards This equipment complies with all requirements of the Medical Device Regulation EU 2017 745 Medical...

Страница 13: ...rs be liable for any indirect special incidental or consequential damages arising from the use of or inability to use the product including without limitation damages for loss of goodwill work and pro...

Страница 14: ...EN109 1560750 44 Copyright Enraf Nonius B V Vareseweg 127 3047 AT Rotterdam The Netherlands Tel 31 0 10 20 30 600 info enraf nonius nl...

Отзывы: