Adjustment of the sections
When adjusting the sections,
squeezing-danger can arise at contact
points of the section. Keep enough
distance between you or your clients
and these points.
9
Cleaning, maintenance and troubleshooting
Cleaning
Use a clean couch cover / towel / terry cloth / paper cover for every individual. This is required
to guarantee a hygienic use.
Do not use agents that contain alcohol or other aggressive cleaners.
•
The couch must be cleaned before it is first used and before it is disinfected.
•
Clean the couch with a damp cloth.
•
Use a mild detergent to clean the surface.
•
Remove soap residue with a damp cloth and clean water.
Disinfection
•
Clean the couch as described above before proceeding to disinfect the couch.
•
Follow the disinfectant's instructions for use.
•
Never use a mixture of disinfectants or a disinfectant and soap mixture.
•
Preferably use a cloth to apply the agent to the imitation leather.
•
Avoid water accumulation in the hinge points of the metal frame. (In the worst case, exposing
the hinges to water in the long-term can result in rust formation)
•
The disinfectants listed on the information sheet (article 0167329), are tested and safe for use
on the imitation leather of the treatment table.
For the most current version of the list see
www.enraf-nonius.com/downloads.
Remark:
Since we have no control over the production and raw materials used for the disinfectants, we
are unable to give a general guarantee that these agents will not damage the upholstery.