  When adjusting this couch, keep hands and feet away from the couch for at least 20 cm, 

especially on places that has been marked by these stickers. 

  When adjusting the sections, squeezing-danger can arise at contact points of the section. 

Keep enough distance between you or your clients and these points. 

  Head section:

 

is not suited to sit upon and must be used to support the head only. 

  Always use the original Enraf-Nonius accessories that are meant for this product. 
  Use a clean couch cover / towel / terry cloth / paper cover for every individual. This is required 

to guarantee a hygienic use. 

  Do not use cleaning agents that contain alcohol or other aggressive cleaners. 
  The Manumed ST should be checked once a year by authorized personnel of Enraf-Nonius or 

by its authorized distributor. 
The manufacturer will not be held responsible for the results of maintenance or repairs by 
unauthorized persons! 

 
 

 

7

 

Operation 

  Use both hands to change the couch settings: one hand for unlocking and the other to adjust 

the section. 

 

Never adjust a section when it supports the full weight of the patient!

 Each section of 

this Manumed ST is supported by gas springs. As well as easing operation, they offer full 
support of the sections. However, these gas springs are not strong enough to lift a section 
while it supports the full weight of the patient. 

 

A maximum patient’s weight of 

150 kg

 is allowed when the height of the couch is being 

adjusted. 

  Be aware of the duty cycle of the motor. The motor is not designed for continuous operation. 

For every 2 minutes the motor is running, it needs 18 minutes to cool down. If you do use the 
motor considerably longer without allowing it to cool down sufficiently, a thermal cut shuts 
off the power and protects the motor against overheating. In that case, let the motor cool 
down until it functions again. 

  The couch is not anti-static. 

 

The operation of your Manumed ST treatment couch is simple and safe.  
The illustrations show how the various adjustments are done. 

 

Moving the couch 
 

The couch can be moved by using the foot pedal to 
lower the wheels (optional).

 

NOTE: Never move the couch with a person or load on 
it. Move the couch only on a smooth surface. The couch 
is not designed to move across thresholds. 

 

 

For safety reasons, the wheels of the unit should 
be  retracted  before  the  beginning  of  a 
treatment. 

 

 

 

Содержание Manumed ST

Страница 1: ...Manumed ST EN109 1560750 44 APRIL 4 2022 Instructions for use...

Страница 2: ...3 2 Symbols 3 3 Package contents 4 4 Installation 5 5 Intended use and intended user 5 6 Precautionary instructions 6 7 Operation 7 8 Safety instructions 9 9 Cleaning maintenance and troubleshooting...

Страница 3: ...o Manumed ST 2 section head section Manumed ST 2 section back section Manumed ST 3 section Manumed ST accessories 2 Symbols Symbol used Description Follow the instructions in the Instructions for Use...

Страница 4: ...d water by mechanical casings and electrical enclosures Double Insulation Type B applied part complying with the specified requirements to provide protection against electric shock particularly regard...

Страница 5: ...d ST should be placed on a flat surface Turn the adjustable foot until it touches the ground and the couch is standing stable For correct assembly and safe use of possible optional accessories one mus...

Страница 6: ...nt to be used by competent personnel in physiotherapy rehabilitation or adjacent areas Avoid any use of the equipment by unauthorized personnel When used by children supervision by a knowledgeable par...

Страница 7: ...upports the full weight of the patient Each section of this Manumed ST is supported by gas springs As well as easing operation they offer full support of the sections However these gas springs are not...

Страница 8: ...ed underneath the framework with your foot the required height of the couch can be set At the head section of the couch the height can be adjusted by turning the handle to the left or to the right Nev...

Страница 9: ...eside the couch is set in flexion position combined with the lowest possible position Underneath the head section between the head section and floor Danger of tilting Do not sit on the head and leg se...

Страница 10: ...clean water Disinfection Clean the couch as described above before proceeding to disinfect the couch Follow the disinfectant s instructions for use Never use a mixture of disinfectants or a disinfecta...

Страница 11: ...service life This device will remain suitable for its intended use as long as it is subjected to yearly maintenance by a qualified service engineer as described in the service manual and the service...

Страница 12: ...EN ISO 10993 5 DIN EN ISO 10993 10 Technical modifications reserved Safety and performance standards This equipment complies with all requirements of the Medical Device Regulation EU 2017 745 Medical...

Страница 13: ...rs be liable for any indirect special incidental or consequential damages arising from the use of or inability to use the product including without limitation damages for loss of goodwill work and pro...

Страница 14: ...EN109 1560750 44 Copyright Enraf Nonius B V Vareseweg 127 3047 AT Rotterdam The Netherlands Tel 31 0 10 20 30 600 info enraf nonius nl...

Отзывы: