Installation and Operation Manual – AQUATEC 1000
Installation
Doc.-ID: AQUA_16012020_EN
9
2.2
Installation of Probe Signal Cable FEP-0001
Abide by the maximum cable length (max. 150m)
Note the minimum bending radius.
FEP-0001
R
min
= 96 mm
Temp. during installation
Temp. during operation
Cross the probe signal cable
(FEP-0001) at right angle
to power supply cables
Figure 5 -
Probe cable FEP-0001
Probe Cable FEP-0001 (FEP-0004 armored version)
No. Function
Diameter
Colors
Additional
info.
Measuring cell
2 x 0,75 mm
2
white-brown / brown With shield
Thermocouple
2 x 0,75 mm
2
green / white
With shield
Probe heating
3 x 1,5 mm
2
black / blue / green-
yellow
Solenoid valve
2 x 0,75 mm
2
grey / grey-blue
Caution
The shield of the probe cable must only be connected at the electronic housing at the PE terminal. Under no
circumstance should the shield also be connected at the probe.
-5 °C
Min.
+50 °C
Max.
-40 °C
Min.
+90 °C
Max.