Service and Maintenance
Installation and Operation Manual – AQUATEC 1000
32
Doc.-ID: AQUA_16012020_EN
5.5.2
Cement in new Filter
Mix the supplied glue
with water according to the
instructions. Spread the glue into the grooves of the seat
and onto the filter part which comes into contact with the
seat. Push - turn the filter
into the seat. Evenly
smooth the glue between the filter and seat and remove
all excess glue.
Figure 32
- Cement new filter in
Note
The cement dries at room temperature within 24 hours. Using the cement supplied by ENOTEC, all types of ceramic,
basalt and sintered metal filters supplied by can be inserted and fixated.
5.6
Replacing the probe
Warning Hot Surface
The probe may only be removed with heat-insulated gloves.
Before removing the probe, always switch off the supply voltage to the electronic system.
1. Disconnect the wires at the probe connection box.
2. Loosen the bolts connecting the probe to the counter flange and remove the probe
3. Insert the replacement probe using a new gasket. Consider the flue gas direction and adjust the V-Shield (filter head)
accordingly. See chapter
2.9 – Adjusting the V-Shield.
4. Tighten the flange bolts and reconnect the wiring at the connection box.
5. Switch on the power to the analyzer and wait for the probe to reach its setpoint temperature.
6. Carry out a 2-point calibration (under process conditions)