Installation and Operation Manual – AQUATEC 1000
Status Messages
Doc.-ID: AQUA_16012020_EN
39
Error Messages
(cont.)
Error
Message
Relay
contact
O
2
signal
output
Description
O
2
sensor
offset too
high
System
error,
open
The error can only occur during an O
2
sensor calibration. An error reset can be carried
out by the user. A successful re-calibration also resets the error.
Possible causes:
defective false test gas (not applicable with test air), test gas flow too
low, O
2
sensor defective.
O
2
sensor
slope too low
System
error,
open
The error can only occur only during an O
2
sensor calibration. An error reset can be
carried out by the user. A successful re-calibration also resets the error.
Possible causes:
Incorrect calibration gas, test gas flow too low, probe filter damaged,
filter head missing, faulty O
2
sensor.
O
2
sensor
slope too high
System
error,
open
The error can only occur only during an O
2
sensor calibration. An error reset can be
carried out by the user. A successful re-calibration also resets the error.
Possible causes:
Incorrect calibration gas, O
2
sensor defective.
O
2
sensor
signal
instable
System
error,
open
The error can only occur only during an O
2
sensor calibration. An error reset can be
carried out by the user. A successful re-calibration also resets the error.
Possible causes:
test gas flow too low, probe filter is damaged, process pressure
fluctuations too high.
mA input for
process
pressure
System
error,
open
Circuit open or pressure transmitter signal less than 3.6 mA / greater than 20.4 mA
Error
REMOTE
module
System
error,
open
Indicates a hardware error of the REMOTE
module.
Possible cause:
the REMOTE
module is defective.
Unlisted errors: Other errors cannot be remedied by the customer. Please contact a service point
6.2
Alarm Messages
Error Message
Relay
contact
Description
Reference air flow too
low
Possible Cause:
reference air flow set incorrectly, instrument air supply is inadequate,
reference air pump is faulty.
Reference air flow too
high
Possible Cause:
reference air flow is set incorrectly.
O
2
limit alarm 1
O
2
limit
alarm 1,
open
Possible Cause:
Over or under range of the O
2
limit alarm.
O
2
limit alarm 2
O
2
limit
alarm 2,
open
Possible Cause:
Over or under range of the O
2
limit alarm.
Electronic temp. too low
Possible Cause:
The ambient temperature of the transmitter is lower than the specified
lower limit. The specified measurement tolerances are no longer guaranteed.
Electronic temp. too high
Possible Cause:
The ambient temperature of the transmitter is higher than the specified
upper limit. The specified measurement tolerances are no longer guaranteed.
Clock battery low
The alarm can only be reset by the user after replacing the clock battery (Lithium type
2032). As long as the system is connected to AC power, the alarm has no impact. Only
after restarting system will the time / date be incorrect. A possible timed automatic
calibration can no longer work correctly.
Unlisted alarm messages: Other messages cannot be remedied by the customer. Please contact a service point
6.3
Maintenance Messages
Maintenance message
Relay
contact
Description
Measured value(s) held
Service,
closed
When
measured values held
is set to on, the determined O
2
-mA output remains saved
for a duration before a calibration.